FDA Adverse Event Malfunction Summary report: N

SOLUTION ONLY GRAVITY IV SET, VENTED, CLAVE X1, 190 CM

MDR report key: 10539456 · Received September 16, 2020

Report

Report Number
3005515211-2020-00003
Event Type
Malfunction
Date Received
September 16, 2020
Date of Event
August 20, 2020
Report Date
August 20, 2020
Manufacturer
AMSINO MEDICAL (SHANGHAI) CO., LTD.
Product Code
FMG
UDI-DI
00(01)(17)(10)
PMA / PMN Number
K964435
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H11: SUBSEQUENT TO THE INITIAL SUBMISSION, IT WAS FOUND THAT LIST NUMBER 32.828C, OR SIMILAR DEVICE, IS NOT MARKETED IN THE U.S. BY ICU MEDICAL, INC., THEREFORE NOT COVERED BY K964435. UPON FURTHER REVIEW, IT WAS DETERMINED THAT THE MEDICAL DEVICE REPORT WAS INADVERTENTLY SUBMITTED.

Additional Manufacturer Narrative · 1

MANUFACTURING PLANT ADDRESS: (B)(4). A SAMPLE WAS RECEIVED FOR EVALUATION. THE INVESTIGATION IS PENDING.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A GRAVITY ADMINISTRATION SET, VENTED WITH CLAVE THAT THE ROLLER CLAMPS WERE TOO LOOSE TO BE HELD AT THE MIDDLE POSITION AND SOME POSITIONS WHERE THEY PREFERRED FOR THE DRIP RATE. THE CLAMPS WOULD DISPOSITION BACK TO THE UNCLAMPED STATUS, WHICH AFFECTS THEIR PRACTICE OF APPLYING THE IV SET ON THE PATIENTS AS THEY ALWAYS SET THE DRIP RATE SAY Q6, Q8, AND THEY WOULD COME BACK 6HRS OR 8HRS LATER FOR CHECKING. DUE TO THE LOOSE ROLLER CLAMP, THE SALINE TO BE DRIPPED FOR THE PATIENT RAN EMPTY EARLIER, 2-3 HOURS BEFORE THEY COME BACK FOR CHECKING. THE PATIENT WAS CONNECTED WITH AN EMPTY NS BOTTLE/BAG FOR 2-3 HOURS, BUT NO HARM AND NO MEDICAL INTERVENTION WERE REPORTED. THIS REPORT CAPTURES THE THIRD OF FIVE EVENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1008748 SOLUTION ONLY GRAVITY IV SET, VENTED, CLAVE X1, 190 CM STOPCOCK, I.V. SET FMG AMSINO MEDICAL (SHANGHAI) CO., LTD. UNKNOWN 00(01)(17)(10)

Patients

Seq Age Sex Outcome Treatment
1