FDA Adverse Event Injury Summary report: N

DEEP BRAIN STIMULATION LEAD

MDR report key: 1053643 · Received May 23, 2008

Report

Report Number
2182207-2008-02766
Event Type
Injury
Date Received
May 23, 2008
Report Date
April 23, 2008
Manufacturer
MEDTRONIC NEUROMODULATION
Product Code
MHY
PMA / PMN Number
P960009
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

JOURNAL REFERENCE: PENA E, PASTOR J, HERNANDO V, ET AL. SKIN EROSION OVER IMPLANTS IN DEEP BRAIN STIMULATION PATIENTS. STEREOTACT FUNCT NEUROSURG 2008;86(2):120-126. WE ANALYZED FACTORS ASSOCIATED WITH SKIN EROSION IN 55 PATIENTS TREATED WITH DEEP BRAIN STIMULATION (KINETRA OR SOLETRA) FOR PARKINSON'S (PD) OR OTHER DISEASES. SOME PATIENTS DEVELOPED EROSION, ALL OF WHOM WERE PD PATIENTS WHO HAD BEEN FITTED WITH A KINETRA DEVICE. EROSIONS MAY BE DUE TO AN INCREASED PRESSURE OVER THE SKIN RESULTING FROM THE LARGER SIZE AND WEIGHT OF THE KINETRA DEVICE. REPORTABLE EVENT: A FEMALE PT WITH PARKINSON'S DISEASE WAS IMPLANTED WITH A KINETRA AND BILATERAL LEADS. AT 14 MONTHS POST, THE PT DEVELOPED EROSION IN THE SKIN OVER RIGHT LEAD AT THE BURR HOLE, ALONG WITH INFECTION. RECURRENCE OF EROSION DID OCCUR; THE RIGHT LEAD AND EXTENSION CABLE WERE REMOVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DEEP BRAIN STIMULATION LEAD MHY MEDTRONIC NEUROMODULATION 3389 UNK

Patients

Seq Age Sex Outcome Treatment
1 71 YR Other| R NEURO STIMULATOR KINETRA MODEL 7428 N=1| LEAD MODEL 3389 N=1| EXTENSION MODEL 7482 N=2| PROGRAMMER MODEL UNK N=1