FDA Adverse Event Injury Summary report: N

LOW PROFILE EXTENSION

MDR report key: 1053634 · Received May 23, 2008

Report

Report Number
2182207-2008-02786
Event Type
Injury
Date Received
May 23, 2008
Report Date
April 23, 2008
Manufacturer
MEDTRONIC NEUROMODULATION
Product Code
MHY
PMA / PMN Number
P960009
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

JOURNAL REFERENCE: PENA E, PASTOR J, HERNANDO V, ET AL. SKIN EROSION OVER IMPLANTS IN DEEP BRAIN STIMULATION PTS. STEREOTACT FUNCT NEUROSURG 2008;86(2):120-126. WE ANALYZED FACTORS ASSOCIATED WITH SKIN EROSION PTS TREATED WITH DEEP BRAIN STIMULATION (KINETRA OR SOLETRA) FOR PARKINSON'S (PD) OR OTHER DISEASES. SOME PTS DEVELOPED EROSION, ALL OF WHOM WERE PD PTS WHO HAD BEEN FITTED WITH A KINETRA DEVICE. EROSION MAY BE DUE TO AN INCREASED PRESSURE OVER THE SKIN RESULTING FROM THE LARGER SIZE AND WEIGHT OF THE KINETRA DEVICE. REPORTABLE EVENT: A MALE PT WITH PARKINSON'S DISEASE WAS IMPLANTED WITH A KINETRA AND A LEFT UNILATERAL LEAD. AT 11 MONTHS POST, THE PT DEVELOPED EROSION IN THE SKIN ALONG THE EXTENSION CABLE TUNNEL PATHWAY IN THE NECK. RECURRENCE OF EROSION DID OCCUR; THE LEFT LEAD WAS REMOVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LOW PROFILE EXTENSION MHY MEDTRONIC NEUROMODULATION 7482 NA

Patients

Seq Age Sex Outcome Treatment
1 57 YR Other| R LEAD MODEL 3889 N=1| PROGRAMMER MODEL UNK N=1| NEURO STIMULATOR KINETRA MODEL 7428 N=1