FDA Adverse Event Injury Summary report: N

LOW PROFILE EXTENSION

MDR report key: 1053632 · Received May 23, 2008

Report

Report Number
2182207-2008-02784
Event Type
Injury
Date Received
May 23, 2008
Report Date
April 23, 2008
Manufacturer
MEDTRONIC NEUROMODULATION
Product Code
MHY
PMA / PMN Number
P960009
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

JOURNAL REFERENCE: PENA E, PASTOR J, HERNANDO V, ET AL. SKIN EROSION OVER IMPLANTS IN DEEP BRAIN STIMULATION PTS. STEREOTACT FUNCT NEUROSURG 2008;86(2):120-126. WE ANALYZED FACTORS ASSOCIATED WITH SKIN EROSION IN 55 PTS TREATED WITH DEEP BRAIN STIMULATION (KINETRA OR SOLETRA) FOR PARKINSON'S (PD) OR OTHER DISEASES. NINE OF 55 PTS DEVELOPED EROSION, ALL OF WHOM WERE PD PTS WHO HAD BEEN FITTED WITH A KINETRA DEVICE. EROSION MAY BE DUE TO AN INCREASED PRESSURE OVER THE SKIN RESULTING FROM THE LARGER SIZE AND WEIGHT OF THE KINETRA DEVICE. REPORTABLE EVENT: A 61 YR OLD PT WITH PARKINSON'S DISEASE WAS IMPLANTED WITH A KINETRA AND BILATERAL LEADS. AT 3 MONTHS POST, THE PT DEVELOPED EROSION IN THE SKIN ALONG THE EXTENSION CABLE TUNNEL PATHWAY IN THE SCALP BEFORE THE CONNECTION, ALONG WITH INFECTION. THE EXTENSION WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LOW PROFILE EXTENSION MHY MEDTRONIC NEUROMODULATION 7482 NA

Patients

Seq Age Sex Outcome Treatment
1 61 YR Other| R PROGRAMMER MODEL UNK N=1| NEURO STIMULATOR KINETRA MODEL 7428 N=1| EXTENSION MODEL 7482 N=1| LEAD MODEL 3389 N=2