FDA Adverse Event Malfunction Summary report: N

THERMOPHORE AUTOMATIC MOIST HEAT PACK

MDR report key: 1053524 · Received May 28, 2008

Report

Report Number
1811605-2008-00079
Event Type
Malfunction
Date Received
May 28, 2008
Date of Event
April 1, 2008
Report Date
April 29, 2008
Manufacturer
BATTLE CREEK EQUIPMENT CO
Product Code
IRT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

IT WAS REPORTED AN INTERNAL ELECTRICAL COMPONENT BURNED. OUR INSPECTION OF THE UNIT REVEALED A BROKEN TERMINAL, RESULTING IN A 1/4" BURN HOLE IN THE FABRIC FOUNDATION OF THE UNIT. SEVERAL INTERNAL COMPONENTS WERE BENT AND BROKEN, INDICATING THAT THE UNIT HAD BEEN FOLDED AND CRUSHED REPEATEDLY - BOTH OF WHICH ARE CLEAR VIOLATIONS OF OUR USER WARNINGS AND INSTRUCTIONS. WE DETERMINED THAT THIS REPORT WAS ATTRIBUTABLE TO MISUSE ON THE PART OF THE USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 THERMOPHORE AUTOMATIC MOIST HEAT PACK ELECTRIC MOIST HEAT PACK IRT BATTLE CREEK EQUIPMENT CO 055

Patients

Seq Age Sex Outcome Treatment
1