FDA Adverse Event
Malfunction
Summary report: N
THERMOPHORE AUTOMATIC MOIST HEAT PACK
MDR report key: 1053524
·
Received May 28, 2008
Report
- Report Number
- 1811605-2008-00079
- Event Type
- Malfunction
- Date Received
- May 28, 2008
- Date of Event
- April 1, 2008
- Report Date
- April 29, 2008
- Manufacturer
- BATTLE CREEK EQUIPMENT CO
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
IT WAS REPORTED AN INTERNAL ELECTRICAL COMPONENT BURNED. OUR INSPECTION OF THE UNIT REVEALED A BROKEN TERMINAL, RESULTING IN A 1/4" BURN HOLE IN THE FABRIC FOUNDATION OF THE UNIT. SEVERAL INTERNAL COMPONENTS WERE BENT AND BROKEN, INDICATING THAT THE UNIT HAD BEEN FOLDED AND CRUSHED REPEATEDLY - BOTH OF WHICH ARE CLEAR VIOLATIONS OF OUR USER WARNINGS AND INSTRUCTIONS. WE DETERMINED THAT THIS REPORT WAS ATTRIBUTABLE TO MISUSE ON THE PART OF THE USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERMOPHORE AUTOMATIC MOIST HEAT PACK | ELECTRIC MOIST HEAT PACK | IRT | BATTLE CREEK EQUIPMENT CO | 055 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |