FDA Adverse Event Malfunction Summary report: N

CONTOUR TEST STRIPS (50)

MDR report key: 1053386 · Received May 22, 2008

Report

Report Number
1826988-2008-00536
Event Type
Malfunction
Date Received
May 22, 2008
Date of Event
April 24, 2008
Report Date
April 24, 2008
Manufacturer
BAYER HEALTHCARE, LLC
Product Code
NBW
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE CUSTOMER STATED THAT SHE TESTED HER BLOOD GLUCOSE USING HER CONTOUR METER AND RECEIVED A READING OF 359 MG/DL. SHE RETESTED USING ANOTHER CONTOUR METER AND RECEIVED A READING OF 156 MG/DL. THE DIFFERENCE BETWEEN THE METER READINGS FALLS IN THE "C" ZONE OF THE PARKES ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS ARE TO BE RETURNED FOR EVAL. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR TEST STRIPS (50) BLOOD GLUCOSE TEST STRIPS NBW BAYER HEALTHCARE, LLC 7097B 7DC3D07

Patients

Seq Age Sex Outcome Treatment
1 UNK