PLANNED PARENTHOOD URINE CASSETTE
Report
- Report Number
- 3005689981-2008-00024
- Event Type
- Malfunction
- Date Received
- May 28, 2008
- Date of Event
- February 18, 2008
- Report Date
- May 28, 2008
- Manufacturer
- INNOVACON, INC.
- Product Code
- LCX
- PMA / PMN Number
- 993317
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
RETENTION DEVICE TESTING. SUMMARY OF RESULTS: THE RETENTION DEVICES MEET QC SPEC. CORRECT POSITIVE RESULTS WERE OBSERVED WHEN TESTED WITH 25MIU/ML HCG URINE CONTROL. THE 100MIU/ML AND 248.57IU/ML HCG URINE CONTROL YIELDED CLEARLY POSITIVE RESULTS AT 3 MINUTES READING TIME. CONCLUSION: THE RETENTION DEVICES MEET QC SPEC. NO FALSE NEGATIVE RESULT WAS OBSERVED. SO THIS COMPLAINT IS NOT CONFIRMED.
A PA CAME IN FOR IUD INSERTION AND RAPID URINE HCG TEST WAS PERFORMED AND RESULT "NEGATIVE". PT CLAIMS LMP WAS 2008. OFFICE INSERTED IUD. PT RETURNED TO OFFICE, REASON UNK, THE FOLLOWING MONTH, (TIME UNK) AND RAPID URINE HCG TEST WAS "POSITIVE" THIS TIME. AN ULTRASOUND WAS PERFORMED AND SHOWED A 6 WEEK GESTATION ALONG WITH HER IUD. PT RETURNED ONE WEEK LATER, FOR PREGNANCY COUNSELING AND HAD ANOTHER ULTRASOUND WHICH SHOWED THE SAME AS THE FIRST. NO OTHER INFO WAS AVAILABLE ON THESE 2 PTS.
IN 2008, (TIME NOT GIVEN), A PT CAME IN FOR AN IUD INSERTION SO A RAPID URINE HCG TEST WAS PERFORMED AND THE RESULT WAS "NEGATIVE". HOWEVER, THE IUD WAS DROPPED AND BROKE SO AN IDU WAS NOT INSERTED THAT DAY. THE PT CAME BACK TO THE OFFICE THE FOLLOWING MONTH, (TIME NOT GIVEN) FOR IUD INSERTION AND ANOTHER RAPID URINE HCG TEST WAS DONE AND IT SHOWED "NEGATIVE" RESULTS AGAIN SO AN IUD WAS INSERTED. PT 2 CLAIMS HER LMP WAS 2008. THE FOLLOWING MONTH, (TIME NOT GIVEN) THIS SAME PT CAME BACK BECAUSE SHE "FELT PREGNANT" AND HAD DONE A HOME PREGNANCY TEST WHICH WAS POSITIVE. A RAPID URINE HCG TEST WAS PERFORMED AND THIS TIME THE RESULT WAS "POSITIVE". AN ULTRASOUND WAS PERFORMED WHICH ALSO SHOWED A 6 WEEK INTRAUTERINE GESTATION. NO OTHER INFO WAS AVAILABLE ON THESE 2 PTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PLANNED PARENTHOOD URINE CASSETTE | ONE STEP PREGNANCY TEST (URINE) | LCX | INNOVACON, INC. | FHC-102-KPP40 | HCG7060082 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 30 YR | |||
| 2 | 21 YR |