FDA Adverse Event Malfunction Summary report: N

FOOTED ATTACHMENT 2.4MM LEGEND, AF02 (B-1)

MDR report key: 1053236 · Received May 28, 2008

Report

Report Number
1625507-2008-00037
Event Type
Malfunction
Date Received
May 28, 2008
Date of Event
April 18, 2008
Report Date
May 1, 2008
Manufacturer
MEDTRONIC POWERED SURGICAL SOLUTIONS
Product Code
HBB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

COMMENTS: THE DEVICE HAS NOT BEEN RETURNED FOR EVAL. OUR RECOMMENDATION FOR FACTORY SERVICE IS BASED ON THE FACILITY'S USAGE, HOWEVER, IT SHOULD NOT EXCEED 24 MONTHS. THIS DEVICE HAS BEEN IN USE FOR 30 MINUTES WITHOUT ANY RECORD OF FACTORY SERVICE. THE USER MANUAL STATES "THE ATTACHMENT SHOULD NOT BE USED IF ANY PART OF THE ATTACHMENT APPEARS TO BE BENT, LOOSE, MISSING, OR DAMAGED. EXCESSIVE PRESSURE OR IMPROPER HANDLING, SUCH AS BENDING OR PRYING, OF THE ATTACHMENT OR DISSECTING TOOL MAY CAUSE INJURY TO THE PT, OPERATOR AND/OR OPERATING ROOM STAFF."

Description of Event or Problem · 1

REPORT RECEIVED STATING THAT DURING A CRANIOTOMY PROCEDURE, THE TIP OF THE FOOTED ATTACHMENT CAME OFF. THE PIECE FELL INTO THE WOUND SITE BUT WAS QUICKLY RETRIEVED. THERE WERE NO FURTHER COMPLICATIONS OBSERVED AND THE CASE WAS SUCCESSFULLY COMPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FOOTED ATTACHMENT 2.4MM LEGEND, AF02 (B-1) PNEUMATIC SURG DRILL ATTACHMENT LEGEND HBB MEDTRONIC POWERED SURGICAL SOLUTIONS NA NA

Patients

Seq Age Sex Outcome Treatment
1