HARMONIC ACE 36 CM
Report
- Report Number
- 1527736-2008-02825
- Event Type
- Injury
- Date Received
- May 8, 2008
- Date of Event
- April 21, 2008
- Report Date
- April 22, 2008
- Manufacturer
- ETHICON ENDO SURGERY, INC. (CINCINNATI)
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER
Narratives
EVAL SUMMARY THE ANALYSIS RESULTS CONFIRMED THAT THE DEVICE A WAS RETURNED WITH THE TISSUE PAD MISSING. NO NOISES WERE NOTED DURING TESTING ON A GENERATOR. AS THE TISSUE PAD WAS NOT RETURNED, FURTHER EVAL COULD NOT BE PERFORMED. HOWEVER, THE CAUSE OF THIS TYPE OF DAMAGE IS CURRENTLY BEING INVESTIGATED; A CORRECTIVE AND PREVENTIVE ACTION HAS BEEN INITIATED TO ADDRESS THE ROOT CAUSE OF IT. EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED DURING THIS INSPECTION AND TESTING. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MFG PROCESS. THE ANALYSIS RESULTS FOUND THAT THE DEVICE B WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED; NO ANOMALIES WERE FOUND WITH THE TISSUE PAD. WHEN A BLADE HAS BEEN COMPROMISED, SCRATCHED OR CRACKED, THE TITANIUM METAL IS FATIGUED WHEN CONTINUALLY ENERGIZED AND THIS RESULTS IN THE BLADE FURTHER CRACKING AND POSSIBLY BREAKING OFF. DURING THE ANALYSIS PROCESS, THE BLADE WAS TESTED FOR SCRATCHES AND CRACKS AND THE BLADE BROKE OFF. DUE TO THE DAMAGE TO THE BLADE, IT WAS CONFIRMED THAT THE DEVICE WAS NON-FUNCTIONAL. THE DEVICE WAS TESTED WITH A GENERATOR AND AN ERROR CODE 5 WAS DISPLAYED; NO NOISE ISSUES WERE NOTED. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. FOR THIS REASON THE FOLLOWING STATEMENTS WERE INCLUDED IN THE INSTRUCTIONS FOR USE: "AVOID ACCIDENTAL CONTACT WITH ALL METAL OR PLASTIC INSTRUMENTS OR OBJECTS WHEN THE INSTRUMENT IS ACTIVATED. CONTACT WITH STAPLES, CLIPS OR OTHER INSTRUMENTS WHILE THE INSTRUMENT IS ACTIVATED MAY RESULT IN CRACKED OR BROKEN BLADES, WHICH MAY BE IDENTIFIED BY GENERATOR SOLID TONE OR INSTRUMENT ERROR." EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED DURING THIS INSPECTION AND TESTING. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MFG PROCESS.
IT WAS REPORTED THAT DURING A NEPHRECTOMY PROCEDURE, THE TISSUE PAD CAME OFF THE DEVICE AND COULD NOT BE FOUND. A SECOND INSTRUMENT WAS A DRILLING SOUND. NO FURTHER INFO PROVIDED ON HOW CASE COMPLETED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC ACE 36 CM | LFL | ETHICON ENDO SURGERY, INC. (CINCINNATI) | NA | E4KD59 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HANDPIECE| GENERATOR |