FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 1052845 · Received May 27, 2008

Report

Report Number
3004464228-2008-00076
Event Type
Malfunction
Date Received
May 27, 2008
Date of Event
April 27, 2008
Report Date
April 27, 2008
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K042792
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE EVAL INDICATES THAT A RETAINER ALLOWED A COMPONENT TO MOVE RESULTING IN DAMAGE WHICH PREVENTED THE PLUNGER FROM ADVANCING. THE USER WOULD HAVE BEEN AWARE OF A PROBLEM DURING THE FILL PROCESS. THERE IS RESISTANCE FELT IN FILLING THE POD WITH INSULIN AND A DISTINCT "CRACKLING" NOISE RESULTING FROM STRIPPING THE THREADS OF THE COMPONENT WHEN OVER-PRESSURISED. THE USER IS INSTRUCTED IN THE USER GUIDE TO FREQUENTLY MONITOR THEIR BG LEVELS. BY FOLLOWING THESE RECOMMENDATIONS, THE USER WOULD BECOME AWARE OF HIGH BG LEVELS AND COULD USE ANOTHER DEVICE OR BACKUP THERAPY IF REQUIRED.

Description of Event or Problem · 1

CUSTOMER CALLED TO REPORT THAT SHE REMOVED POD BECAUSE SHE WAS NOT SURE IF INSULIN WAS BEING DELIVERED. CUSTOMER SAID HER BG WENT OVER 300 AND PLACED ANOTHER POD. NO FURTHER INFO REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 11200 L11649

Patients

Seq Age Sex Outcome Treatment
1 Other