FDA Adverse Event Malfunction Summary report: N

AFFINITY BED

MDR report key: 1052841 · Received May 27, 2008

Report

Report Number
1824206-2008-02581
Event Type
Malfunction
Date Received
May 27, 2008
Date of Event
May 12, 2008
Report Date
May 12, 2008
Manufacturer
HILL-ROM RITTER
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM FIELD INVESTIGATION INDICATED THE BRAKES WERE DIFFICULT TO SET ON THE BED. THE HILL-ROM TECHNICIAN REPLACED THE BRAKE/STEER CASTER, BRAKE BLOCK AND TRANSLINK TO REPAIR THE BED. THE AFFINITY SERVICE MANUAL (MAN013RE) DOCUMENTS A FUNCTION CHECK FOR THE CASTER TIRES AND CENTRAL BRAKE AND STEER AS PART OF PREVENTATIVE MAINTENANCE. THE COMPONENTS SHOULD BE CHECKED FOR CUTS, WEAR, TREAD LIFE, ETC AND THE BRAKES SHOULD BE CHECKED TO ENSURE THE BED WILL NOT MOVE WHEN THE BRAKE IS ACTIVATED. IF THE BED MOVES, INSPECTION FOR WEAR AND ADJUSTMENT MAY BE NECESSARY.

Description of Event or Problem · 1

ALLEGED WHEN THE BRAKE PEDAL IS SET AND THE BED IS LEANED ON, THE BREAK WILL NOT HOLD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY BED BIRTHING BED HDD HILL-ROM RITTER 3600

Patients

Seq Age Sex Outcome Treatment
1