FDA Adverse Event
Malfunction
Summary report: N
PRISMAFLO II S
MDR report key: 10527504
·
Received September 14, 2020
Report
- Report Number
- 10527504
- Event Type
- Malfunction
- Date Received
- September 14, 2020
- Date of Event
- July 17, 2020
- Report Date
- July 20, 2020
- Manufacturer
- STIHLER ELECTRONIC GMBH
- Product Code
- KOC
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- TX
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
PHYSICIAN AND BEDSIDE NURSE ENTERED PATIENT'S ROOM TO DELIVER PATIENT CARE. WHEN THEY MOVED THE PILLOW ON THE BED BY THE PATIENT'S LEG, THEY OBSERVED THE BURNT AREA ON THE BED AND PILLOW. THEY POURED WATER ON THE BED TO EXTINGUISH IT. A WARMER LINE OF THE HEATER (PRISMAFLO II DEVICE) ATTACHED TO THE CVVH PRISMA FLEX MACHINE, THAT REWARMS BLOOD ON THE RETURN TO THE PATIENT, CAUSED A SMALL FIRE IN THE MATTRESS AND PILLOW. THE TUBING LINE WAS IN THE PATIENT BED AND WAS FOUND TO BE IN BETWEEN THE PILLOW AND THE MATTRESS AND BEGUN TO BURN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 995867 | PRISMAFLO II S | ACCESSORIES, BLOOD CIRCUIT, HEMODIALYSIS | KOC | STIHLER ELECTRONIC GMBH | PF2300NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 5110 DA |