FDA Adverse Event Malfunction Summary report: N

CNS-6201A

MDR report key: 10525871 · Received September 11, 2020

Report

Report Number
8030229-2020-00529
Event Type
Malfunction
Date Received
September 11, 2020
Date of Event
August 12, 2020
Report Date
March 1, 2023
Manufacturer
NIHON KOHDEN CORPORATION
Product Code
MHX
UDI-DI
04931921114131
PMA / PMN Number
K102376
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DETAILS OF COMPLAINT: THE BIOMEDICAL ENGINEER (BME) REPORTED THEY HAD A POWER FAILURE, AND THE UPS CONNECTED TO THE CENTRAL NURSE'S STATION (CNS) FAILED, CAUSING THE CNS TO POWER OFF UNEXPECTEDLY. IT WOULD NOT BOOTUP PAST THE INITIAL BOOT SPLASH SCREEN. NIHON KOHDEN TECHNICAL SUPPORT (NK TS) HAD THEM POWER THE CNS OFF, REMOVE THE PRIMARY HDD AND PUT THE SECONDARY HDD INTO THE PRIMARY PORT, THEN BOOT THE CNS UP WITH JUST ONE HDD. THE CNS BOOTED UP AND LAUNCHED INTO THE CNS SOFTWARE SUCCESSFULLY. ONCE THE SOFTWARE WAS UP AND RUNNING, NK TS HAD THEM INSERT THE OTHER HDD INTO PORT 1. THEY VERIFIED THAT BOTH HDDS WERE GOOD AND WERE REBUILDING THE RAID. NO PATIENT HARM OR INJURY WAS REPORTED. INVESTIGATION SUMMARY: SPONTANEOUS SHUTDOWN OR SPONTANEOUS REBOOT EVENTS ARE TRIGGERED BY EITHER A POWER LOSS OR A HARDWARE FAILURE. THE CAUSES OF EACH ARE DISCUSSED BELOW. POWER LOSS WHEN THE CNS EXPERIENCES A POWER LOSS, IT CAN BE CAUSED BY A VARIETY OF EVENTS. THESE EVENTS INCLUDE POWER OUTAGES, GENERATOR TESTS, UNINTERRUPTIBLE POWER SUPPLY (FAILURE), POWER OUTLET FAILURE. THESE ARE ENVIRONMENTAL FACTORS THAT IMPACT DEVICE FUNCTIONALITY. HARDWARE FAILURE HARDWARE FAILURE THAT MAY RESULT IN SPONTANEOUS SHUTDOWN OR REBOOT INCLUDES POWER CORD FAILURE, HARD DRIVE FAILURE, AND VARIOUS OTHER ESSENTIAL COMPONENTS OF A COMPUTER SYSTEM SUCH AS THE COOLING FAN, INTERNAL POWER SUPPLY, MORTARBOARD, CPU, MEMORY, ETC. MOST COMMONLY, HARD DRIVE FAILURES WILL RESULT IN SPONTANEOUS SHUTDOWN OR REBOOT. IN THIS PARTICULAR CASE, THE USER'S FACILITY EXPERIENCED A POWER BUMP WHICH CAUSED ABRUPT POWER LOSS TO THE CNS. THE ABRUPT POWER LOSS CAUSED HARD DRIVE FAILURE, RESULTING IN THE FAILURE TO REBOOT BACK INTO THE CNS APPLICATION.

Description of Event or Problem · 0

THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THEY HAD A POWER FAILURE, AND THE UPS CONNECTED TO THE CENTRAL NURSE'S STATION (CNS) FAILED, CAUSING THE CNS TO POWER OFF UNEXPECTEDLY. IT WOULD NOT BOOTUP PAST THE INITIAL BOOT SPLASH SCREEN.

Additional Manufacturer Narrative · 1

THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THEY HAD A POWER FAILURE, AND THERE WAS A UPS CONNECTED TO THE CENTRAL NURSE'S STATION (CNS), BUT IT FAILED AND CAUSED THE CNS TO POWER OFF UNEXPECTEDLY. THEN THEY TRIED TO POWER ON THE CNS, BUT IT WOULD NOT GET PAST THE INITIAL BOOT SPLASH SCREEN. NIHON KOHDEN TECHNICAL SUPPORT HAD THEM POWER THE CNS OFF, REMOVE THE PRIMARY HDD AND PUT THE SECONDARY HDD INTO THE PRIMARY PORT, AND BOOT THE CNS UP WITH JUST ONE HDD. THE CNS BOOTED UP AND LAUNCHED INTO THE SOFTWARE SUCCESSFULLY. ONCE THE SOFTWARE WAS UP AND RUNNING, TECHNICAL SUPPORT HAD THEM INSERT THE OTHER HDD INTO PORT 1. THEN THEY CHECKED THE INTEL STORAGE MATRIX, AND IT SHOWED THAT BOTH HDD'S WERE GOOD AND IT WAS REBUILDING THE RAID. NO PATIENT HARM REPORTED. NIHON KOHDEN CONTINUES TO INVESTIGATE THE REPORTED EVENT. NIHON KOHDEN WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE. ADDITIONAL DEVICE INFORMATION: CONCOMITANT MEDICAL DEVICE FIELD CONTAINS MISSING INFORMATION (MI), AS ATTEMPTS TO OBTAIN INFORMATION WERE MADE, BUT NOT PROVIDED.

Description of Event or Problem · 1

THE BIOMEDICAL ENGINEER (BME) REPORTED THAT THEY HAD A POWER FAILURE, AND THERE WAS A UPS CONNECTED TO THE CENTRAL NURSE'S STATION (CNS), BUT IT FAILED AND CAUSED THE CNS TO POWER OFF UNEXPECTEDLY. THEN THEY TRIED TO POWER ON THE CNS, BUT IT WOULD NOT GET PAST THE INITIAL BOOT SPLASH SCREEN. NIHON KOHDEN TECHNICAL SUPPORT HAD THEM POWER THE CNS OFF, REMOVE THE PRIMARY HDD AND PUT THE SECONDARY HDD INTO THE PRIMARY PORT, AND BOOT THE CNS UP WITH JUST ONE HDD. THE CNS BOOTED UP AND LAUNCHED INTO THE SOFTWARE SUCCESSFULLY. ONCE THE SOFTWARE WAS UP AND RUNNING, TECHNICAL SUPPORT HAD THEM INSERT THE OTHER HDD INTO PORT 1. THEN THEY CHECKED THE INTEL STORAGE MATRIX, AND IT SHOWED THAT BOTH HDD'S WERE GOOD AND IT WAS REBUILDING THE RAID. NO PATIENT HARM REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
989803 CNS-6201A CENTRAL MONITOR SYSTEM MHX NIHON KOHDEN CORPORATION CNS-6201A NA 04931921114131

Patients

Seq Age Sex Outcome Treatment
1 Unknown MI.| NI.