FDA Adverse Event Malfunction Summary report: N

INFANT BREATHING CIRCUIT

MDR report key: 1052578 · Received May 22, 2008

Report

Report Number
9611451-2008-00310
Event Type
Malfunction
Date Received
May 22, 2008
Date of Event
April 11, 2008
Report Date
April 22, 2008
Manufacturer
FISHER & PAYKEL HEALTHCARE, LTD.
Product Code
BZE
PMA / PMN Number
K034026
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED TO THE MFR. AN INVESTIGATION WAS CARRIED OUT BASED ON THE EVENT DESCRIPTION AND PREVIOUS EXPERIENCE WITH SIMILAR COMPLAINTS. RESULTS: WE WERE UNABLE TO CARRY OUT A LOT CHECK AS NO LOT INFO WAS PROVIDED WITH THIS COMPLAINT. THE DUCKBILL SUCTION PORT CONTAINS A ONE WAY DUCKBILL VALVE AND IS POSITIONED AT THE ELBOW OF THE Y-PIECE. THE FACT THAT A LEAK WAS DETECT IN THE RT236 BREATHING CIRCUIT ONLY UNTIL THE OPERATOR OCCLUDED THE DUCKBILL SUGGESTS THERE WAS A LEAK IN THE DUCKBILL VALVE. CONCLUSION: ALL BREATHING CIRCUITS ARE PRESSURE TESTED DURING PRODUCTION. THE TEST HAS A SMALL ACCEPTABLE PRESSURE DROP AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS THAT A LEAK IN THE DUCKBILL DEVELOPED AFTER PRODUCTION. WE HAVE AN OPEN CAPA ITEM TO ADDRESS SUCH COMPLAINTS AND TO PUT MEASURES IN PLACE TO REDUCE AND/OR PREVENT RECURRENCE. OUR MONITORING AND TRENDING FOR THIS TYPE OF EVENT SHOWS A RATE OF OCCURRENCE WORLDWIDE FOR THE LAST YEAR OF 0.0037%.

Description of Event or Problem · 1

A HOSP REPORTED THROUGH OUR DISTRIBUTOR THAT A LEAK WAS DETECTED DURING SETUP FOR A QUANTITY OF 2 INFANT BREATHING CIRCUITS. IT WAS REPORTED THE CIRCUIT PASSED THE LEAK TEST WHEN THE OPERATOR OCCLUDED THE DUCK BILL SUCTION PORT. THE CIRCUITS WERE THEN USED ON THE PT WITH NO APPARENT PROBLEM. NO PT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFANT BREATHING CIRCUIT BZE FISHER & PAYKEL HEALTHCARE, LTD. RT236

Patients

Seq Age Sex Outcome Treatment
1 SERVO I VENTILATOR