FDA Adverse Event Malfunction Summary report: N

CC IRON/MAGNESIUM CALIBRATOR

MDR report key: 1052459 · Received May 22, 2008

Report

Report Number
2018433-2008-00105
Event Type
Malfunction
Date Received
May 22, 2008
Date of Event
February 21, 2008
Report Date
April 28, 2008
Manufacturer
ABBOTT MANUFACTURING, INC
Product Code
JIX
PMA / PMN Number
K981706
Removal / Correction Number
2018433-05/05/08-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IT WAS DETERMINED AFTER COMPLETION OF IN-HOUSE TESTING THAT THE IRON/MAGNESIUM CALIBRATOR LOT 54187M200 VALUE ASSIGNMENT SHEET, WAS INCORRECTLY ASSIGNED BY THE OEM WITH A SET POINT FOR CAL 2 AT 3.2 MEQ/L INSTEAD OF 0.8 MEQ/L. AN ASSESSMENT TO PT IMPACT WAS PERFORMED AND VALUES ABOVE 0.8 MEQ/L WERE FOUND TO BE 17.24% LOWER THAN THEY SHOULD BE. THIS IS AN INITIAL REPORT. THE MFR HAS BEEN NOTIFIED AND A FURTHER INVESTIGATION IS CURRENTLY IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT THEY STARTED A NEW LOT OF IRON/MAGNESIUM CALIBRATORS ON THE AEROSET ANALYZER AND THE QUALITY CONTROL SHIFTED LOW FOR MAGNESIUM. THE CUSTOMER STATES THEY RERAN FOUR SAMPLES WITH ORIGINAL RESULTS OBTAINED ON THE PREVIOUS CALIBRATOR LOT. SAMPLE THREE OF FOUR HAD AN ORIGINAL RESULT OF 2.0MEQ/L AND A REPEAT RESULT OF 2.5 MEQ/L. NO RESULTS WERE QUESTIONED BY THE PHYSICIAN. THE CUSTOMER STATED THAT THE INSTRUMENT MAINTENANCE IS UP TO DATE AND THERE IS NO OTHER ASSAY ISSUES. THE ABBOTT CUSTOMER TECHNICAL ADVOCATE SENT A DIFFERENT CALIBRATOR LOT FOR TROUBLESHOOTING. THE CUSTOMER RECEIVED THE NEW CALIBRATOR AND THE QUALITY CONTROL IS BACK IN RANGE. NO IMPACT TO PT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CC IRON/MAGNESIUM CALIBRATOR FOR USE IN THE CALIBRATION OF THE IRON AND MAGNESIUM ASSAYS JIX ABBOTT MANUFACTURING, INC NA 54187M200

Patients

Seq Age Sex Outcome Treatment
1 NI AEROSET ANALYZER, LIST# 9D05-01