CC IRON/MAGNESIUM CALIBRATOR
Report
- Report Number
- 2018433-2008-00103
- Event Type
- Malfunction
- Date Received
- May 22, 2008
- Date of Event
- March 20, 2008
- Report Date
- April 28, 2008
- Manufacturer
- ABBOTT LABORATORIES, INC
- Product Code
- JIX
- PMA / PMN Number
- K981706
- Removal / Correction Number
- 2018433-05/05/08-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
IT WAS DETERMINED AFTER COMPLETION OF IN-HOUSE TESTING THAT THE IRON/MAGNESIUM CALIBRATOR LOT 54187M200 VALUE ASSIGNMENT SHEET, WAS INCORRECTLY ASSIGNED BY THE OEM WITH A SET POINT FOR CAL 2 AT 3.2 MEQ/L INSTEAD OF 0.8 MEQ/L. AN ASSESSMENT TO PT IMPACT WAS PERFORMED AND VALUES ABOVE 0.8 MEQ/L WERE FOUND TO BE 17.24% LOWER THAN THEY SHOULD BE. THIS IS AN INITIAL REPORT. THE MFR HAS BEEN NOTIFIED AND A FURTHER INVESTIGATION IS CURRENTLY IN PROGRESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
THE CUSTOMER STATED THAT AFTER BEGINNING A NEW LOT OF IRON/MAGNESIUM CALIBRATORS ON THE ARCHITECT C8000 ANALYZER, THE QUALITY CONTROL FOR BOTH SERUM AND URINE AND THE PT RESULTS SHIFTED DOWN. THE CUSTOMER ALSO NOTED THE INSTRUMENT WAS RUNNING FINE WITH THE PREVIOUS LOT. THE CUSTOMER PULLED FOUR PREVIOUSLY RUN SAMPLES FOR COMPARISON. ALL FOUR SAMPLE RESULTS SHIFTED DOWN WITH THE NEW CALIBRATOR. SAMPLE ONE OF FOUR HAD AN ORIGINAL RESULT OF 1.4 MEQ/L AND A REPEAT RESULT OF 1.3 MEQ/L. THE ORIGINAL RESULT WAS REPORTED. ALL INSTRUMENT MAINTENANCE IS UP TO DATE. THERE ARE NO KNOWN INSTRUMENT ISSUES OR RECENT PART CHANGES. THE ABBOTT CUSTOMER TECHNICAL ADVOCATE SENT THE CUSTOMER AN ALTERNATE CALIBRATOR LOT FOR TROUBLESHOOTING. A FOLLOW-UP WITH THE CUSTOMER CONFIRMED RECEIPT OF THE NEW LOT OF CALIBRATOR. THE CUSTOMER RECALIBRATED THE INSTRUMENT AND ALL OF THE QUALITY CONTROL RECOVERED IN RANGE. NO IMPACT TO PT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CC IRON/MAGNESIUM CALIBRATOR | FOR USE IN THE CALIBRATION OF THE IRON AND MAGNESIUM ASSAYS | JIX | ABBOTT LABORATORIES, INC | NA | 54187M200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI | ARCHITECT C8000 |