FDA Adverse Event
Malfunction
Summary report: N
ELITE TEST STRIPS (100)
MDR report key: 1052393
·
Received May 23, 2008
Report
- Report Number
- 1826988-2008-00541
- Event Type
- Malfunction
- Date Received
- May 23, 2008
- Date of Event
- April 25, 2008
- Report Date
- April 25, 2008
- Manufacturer
- BAYER HEALTHCARE, LLC
- Product Code
- NBW
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE CUSTOMER STATED THAT SHE TESTED HER BLOOD GLUCOSE AND REC'D A READING OF 400 MG/DL USING HER ELITE METER. SHE RETESTED USING HER CONTOUR METER AND REC'D A READING OF 160 MG/DL. THE DIFFERENCE BETWEEN READINGS FALLS IN THE "C" ZONE OF THE PARKES ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS ARE TO BE RETURNED FOR EVAL. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ELITE TEST STRIPS (100) | BLOOD GLUCOSE TEST STRIPS | NBW | BAYER HEALTHCARE, LLC | 3942A | 7H05MS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |