HI PWR DISPOSBL GRNDING PAD
Report
- Report Number
- 1717344-2008-00224
- Event Type
- Injury
- Date Received
- May 23, 2008
- Date of Event
- May 9, 2008
- Report Date
- May 16, 2008
- Manufacturer
- COVIDIEN LP, (VALLEYLAB)
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE SITE REPORTED, THESE WERE OLD PADS, BUT DID NOT REPORT LOT #. DGPHP SITE BURNS CAN BE THE RESULT OF A NUMBER OF CAUSES INCLUDING PLACEMENT, DURATION OF TREATMENT AND POWER SETTINGS, PATIENT CONDITION, AND POTENTIALLY BY FAILURE OF THE DISPERSIVE ELECTRODE TO PERFORM AS INTENDED. IN THIS CASE, WITHOUT THE RETURN OF THE INCIDENT DISPERSIVE ELECTRODE, WE WERE UNABLE TO DETERMINE IF ANY DEFECT OF THE PRODUCT CAUSED OR CONTRIBUTED TO THE INCIDENT. ONE OF THE CAUSES OF BURNS IS POOR DGPHP PLACEMENT. AS NOTED ABOVE, PROPER PLACEMENT OF THE DGPHP AND ENSURING GOOD CONTACT THROUGHOUT THE PROCEDURE (ESPECIALLY IF THE PT IS REPOSITIONED) ARE CRITICAL COMPONENTS TO AVOIDING BURNS. THIS IS COVERED IN DETAIL IN OUR IFU. FURTHERMORE, WE ARE AWARE THAT SOME CUSTOMERS REUSE PADS EVEN THOUGH THEY ARE CLEARLY LABELED AS SINGLE USE DEVICES. REUSED DISPERSIVE ELECTRODES CAN DRY OUT AND BE A SOURCE FOR BURNS BECAUSE THEY DO NOT CONDUCT ELECTRICAL CURRENT PROPERLY. WE ARE CLOSELY MONITORING THE INCIDENCE RATE OF DISPERSIVE ELECTRODE BURNS. THE RATE OF BURNS FOR VALLEYLAB DGPHP DISPERSIVE ELECTRODES CONTINUES TO BE LOWER THAN THE OVERALL RATE OF RADIO FREQUENCY ABLATION DISPERSIVE ELECTRODE BURNS AS CITED IN CURRENT MEDICAL LITERATURE. CONTINUOUS IMPROVEMENT EFFORTS ARE ON-GOING TO ENSURE THAT CUSTOMERS ARE PROPERLY TRAINED AND OUR IFU CLEARLY ILLUSTRATES THE PROPER PLACEMENT OF THE DISPERSIVE ELECTRODES. WE ALSO REINFORCE THE IMPORTANCE OF NOT REUSING THESE SINGLE USE DEVICES TO THE CUSTOMER WHENEVER THE OPPORTUNITY ARISES.
THE REPORT STATED THAT A RADIO FREQUENCY ABLATION PROCEDURE WAS PERFORMED WITH THE OUTPUT AT 40W FOR FIVE 12-MIN ABLATIONS. TWO HOURS LATER, 3 MORE 12-MIN ABLATIONS WERE DONE. WHEN REMOVING THE PT RETURN ELECTRODES AT THE END OF THE PROCEDURE, A 7X5CM BURN AND A 5X3.5CM BURN WHERE FOUND ON PT'S RIGHT THIGH. THESE WERE REPORTED AS 3RD DEGREE BURNS. THE PT WAS TREATED WITH GEBEN CREAM. THERE WAS NO POSITION CHANGE DURING THE PROCEDURE. THE PT'S LENGTH OF HOSPITAL STAY WAS EXTENDED DUE TO BURN INJURY. THE SITE USED OLD PADS THE CUSTOMER HAD KEPT, NOT THE ONES INCLUDED WITH THE NEEDLE. THE NURSE WHO PUT THE PADS ON THE PT STATES THERE WAS ENOUGH GEL ON THE RIGHT PAD. WHILE THEY FELT THE LEFT PAD HAD LESS GEL, NO BURN WAS FOUND ON THE LEFT THIGH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HI PWR DISPOSBL GRNDING PAD | ELECTROSURGICAL ACCESSORY | GEI | COVIDIEN LP, (VALLEYLAB) | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Hospitalization| O |