FDA Adverse Event Malfunction Summary report: N

OPEN SCREWDRIVER - MODULAR TI

MDR report key: 1052274 · Received May 23, 2008

Report

Report Number
1649384-2008-00277
Event Type
Malfunction
Date Received
May 23, 2008
Date of Event
April 24, 2008
Report Date
May 23, 2008
Manufacturer
ABBOTT SPINE
Product Code
HXX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVENT DID NOT OCCUR IN SURGERY. PRODUCT IS AN INSTRUMENT AND IS NOT IMPLANTABLE. THE RESULTS OF THE DEVICE HISTORY RECORD REVIEW INDICATE THE PART MET SPECIFICATION. VISUAL INSPECTION FOUND THE PRONGS OF THE DRIVER TO BE BROKEN OFF IN THE DIRECTION OF LOOSENING. AN ENGINEERING EVALUATION DETERMINED THAT DAMAGE PRONGS ARE THE RESULT OF ATTEMPTING TO REMOVE HARDWARE WITH THE DRIVER. THE SURGICAL TECHNIQUE WAS UPDATE IN 2007 WITH ENHANCED PRECAUTIONS.

Description of Event or Problem · 1

IN 2008, THE SALES REPRESENTATIVE REPORTED THAT DURING A KIT INSPECTION IT WAS IDENTIFIED THAT THE PRONGS OF THE DRIVER WERE DAMAGED. THE MALFUNCTION HAS RESULTED IN AN ADVERSE EVENT IN THE PAST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OPEN SCREWDRIVER - MODULAR TI INCOMPASS HXX ABBOTT SPINE 25YE

Patients

Seq Age Sex Outcome Treatment
1