FDA Adverse Event Malfunction Summary report: N

PROGRAMMING WAND

MDR report key: 1052244 · Received May 23, 2008

Report

Report Number
1644487-2008-01210
Event Type
Malfunction
Date Received
May 23, 2008
Date of Event
April 18, 2008
Report Date
April 25, 2008
Manufacturer
CYBERONICS, INC.
Product Code
MUZ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE FAILURE IS SUSPECTED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.

Description of Event or Problem · 1

REPORTER INDICATED THAT A VNS WAND FAILED TO PROGRAM A PT AT AN OFFICE VISIT. THE PT HAD PREVIOUSLY BEEN ABLE TO BE INTERROGATED EARLIER AT THE SAME VISIT, BUT THE REPORTER WAS NOW RECEIVING COMMUNICATION ERRORS. A VNS SYSTEMS DIAGNOSTICS TEST DONE EARLIER AT THE SAME VISIT WAS NORMAL. THE VNS WAND, COMPUTER AND FLASHCARD HAVE BEEN RETURNED AND ARE CURRENTLY IN PRODUCT ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING WAND MUZ CYBERONICS, INC. 201

Patients

Seq Age Sex Outcome Treatment
1