FDA Adverse Event
Malfunction
Summary report: N
M SERIES
MDR report key: 1052169
·
Received May 20, 2008
Report
- Report Number
- 1220908-2008-01076
- Event Type
- Malfunction
- Date Received
- May 20, 2008
- Report Date
- May 6, 2008
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE REPORTED MALFUNCTION WAS OBSERVED AND ATTRIBUTED TO FAULTY TRANSISTORS ON THE BRIDGE BOARD. ANALYSIS OF REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE FAILED SELF TEST. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | M SERIES | DEFIBRILLATOR/PACEMAKER | MKJ | ZOLL MEDICAL CORPORATION | MSERIES BI-PHASIC | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |