DIMENSION CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 1226181-2008-00054
- Event Type
- Malfunction
- Date Received
- May 16, 2008
- Date of Event
- April 17, 2008
- Report Date
- April 18, 2008
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- MMI
- PMA / PMN Number
- K944093
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- OTHER
Narratives
NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULTS WAS A SLIPPING HM WASH PUMP DUE TO LACK OF MONTHLY MAINTENANCE BY THE CUSTOMER. THE MALFUNCTION WAS RESOLVED AFTER THE CUSTOMER CORRECTED THE PROBLEM WITH GUIDANCE FROM A SIEMENS HEALTHCARE DIAGNOSTICS INC TECHNICAL SOLUTIONS CENTER REP. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS.
FALSELY ELEVATED TROPONIN I RESULTS WERE OBTAINED ON FOUR PTS' SAMPLES. THE RESULTS WERE REPORTED TO THE PHYSICIANS. THE ORIGINAL SAMPLES WERE RETESTED AND NEGATIVE RESULTS WERE OBTAINED. IT IS UNK IF PT TREATMENT WAS ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULTS. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULTS WAS A SLIPPING HM WASH PUMP DUE TO LACK OF MONTHLY MAINTENANCE BY THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION CLINICAL CHEMISTRY SYSTEM | CLINICAL CHEMISTRY SYSTEM | MMI | SIEMENS HEALTHCARE DIAGNOSTICS INC. | XPANDHM | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |