FDA Adverse Event
Malfunction
Summary report: N
STARCLOSE VASCULAR CLOSURE SYSTEM
MDR report key: 1052073
·
Received May 21, 2008
Report
- Report Number
- 2953144-2008-00925
- Event Type
- Malfunction
- Date Received
- May 21, 2008
- Date of Event
- April 26, 2008
- Report Date
- April 26, 2008
- Manufacturer
- ABBOTT VASCULAR INC.
- Product Code
- MGB
- PMA / PMN Number
- P050007
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. THE LOT NUMBER WAS PROVIDED. A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT DID NOT PRODUCE ANY FINDINGS RELEVANT TO THIS REPORT.
Description of Event or Problem · 1
DEVICE MALFUNCTION: DIFFICULT TO REMOVE. TIME OF MALFUNCTION: DURING VESSEL CLOSURE. SYMPTOMS/AE: NONE. IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE STARCLOSE DEVICE ACHIEVED ARTERIOTOMY CLOSURE OF AN UNSPECIFIED VESSEL AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, AFTER DEPLOYING THE CLIP, THE DEVICE WAS DIFFICULT TO REMOVE. THE DEVICE WAS REMOVED BY USING BOTH "COUNTER-TRACTION AND PULL". THERE WERE NO REPORTED ADVERSE PT EFFECTS. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STARCLOSE VASCULAR CLOSURE SYSTEM | MGB | ABBOTT VASCULAR INC. | NA | 62049-6H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |