FDA Adverse Event
Malfunction
Summary report: N
PRECISION XTRA / OPTIUM
MDR report key: 1052063
·
Received May 20, 2008
Report
- Report Number
- 2954323-2008-01889
- Event Type
- Malfunction
- Date Received
- May 20, 2008
- Date of Event
- March 5, 2008
- Report Date
- May 20, 2008
- Manufacturer
- ABBOTT DIABETES CARE LIMITED UK
- Product Code
- NBW
- Removal / Correction Number
- Z-0112-2008
- Product Problem
- Yes
- Report Source
- Manufacturer report
Narratives
Additional Manufacturer Narrative · 1
PRODUCT TESTING ON THE RETURNED METER CONFIRMED A DISPLAY ISSUE. DAMAGED DISPLAY MAY OCCUR WHEN THE METER IS DROPPED RESULTING IN PAD FAILURES THAT CAUSE SPECIFIC AREAS OF THE METER'S LCD SCREEN TO BE UNREADABLE. CUSTOMERS AND RETAILERS HAVE BEEN INFORMED THROUGH THE ADC FA17AUG2007 LETTER.
Description of Event or Problem · 1
A CUSTOMER REPORTED AN ISSUE WITH THE DISPLAY ON THEIR PRECISION XTRA BLOOD GLUCOSE METER. UPON PRODUCT INVESTIGATION, IT WAS DISCOVERED THE METER EXHIBITED A DISPLAY ISSUE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRECISION XTRA / OPTIUM | BLOOD GLUCOSE MONITORING SYSTEM | NBW | ABBOTT DIABETES CARE LIMITED UK | NI | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI |