FDA Adverse Event Malfunction Summary report: N

GE

MDR report key: 10520409 · Received September 11, 2020

Report

Report Number
9617251-2020-00002
Event Type
Malfunction
Date Received
September 11, 2020
Date of Event
August 17, 2020
Report Date
September 10, 2020
Manufacturer
SEDECAL SA
Product Code
KPR
UDI-DI
08436046002302
PMA / PMN Number
K042876
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

AS MENTIONED IN THE EVENT'S DESCRIPTION, THE CAUSE OF THE PROBLEM IS NOT RELATED, AT ALL, WITH A MALFUNCTION OF THE UNIT BUT A WRONG OR DEFECTIVE MAINTENANCE TASK. THE ASSEMBLY INSTRUCTIONS FOR COLLIMATOR ARE CORRECT AND CLEAR ENOUGH AND THEY ARE INCLUDED IN THE SYSTEM MANUAL AS WELL AS IN THE COLLIMATOR MANUAL. NO FURTHER ACTION IS REQUIRED FROM SEDECAL SIDE, JUST TO REPORT THE EVENT BY MEANS OF A MDR (3500A FORM).

Description of Event or Problem · 1

THE COLLIMATOR FELL OFF TUBEHEAD. THE COLLIMATOR WAS REATTACHED. THERE WAS NO INJURY. THE COLLIMATOR DID DETACH AND FALL ON THE TABLE DUE TO A DEFECTIVE REASSEMBLY AFTER A SERVICE MAINTENANCE TASK AS THE TECHNOLOGIST WAS POSITIONING THE COLLIMATOR DURING A PATIENT CHEST EXAM, THE COLLIMATOR FELL FROM ITS MOUNT ONTO THE TABLE. IMPORTANT INFORMATION: THE COLLIMATOR WAS REMOVED THE DAY BEFORE TO REPLACE TUBE. THE TUBE WAS THEN PUT BACK AS IT WAS NOT THE CAUSE OF THE PROBLEM. THE SET SCREW WAS NOT ALL THE WAY IN. FIELD ENGINEER WENT TO SITE AND SCREWED THE SCREW BACK IN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
990224 GE DIGITAL DIAGNOSTIC X-RAY SYSTEM KPR SEDECAL SA PROTEUS XR/F ET 08436046002302

Patients

Seq Age Sex Outcome Treatment
1