FDA Adverse Event Malfunction Summary report: N

PUMP MMT-1780KR 670G MG MEDI RECERT

MDR report key: 10519503 · Received September 11, 2020

Report

Report Number
2032227-2020-175388
Event Type
Malfunction
Date Received
September 11, 2020
Date of Event
August 30, 2020
Report Date
March 21, 2021
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000365912
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). NEED TO CORRECT PRODUCT CODE FROM OYC TO OZP.

Additional Manufacturer Narrative · 1

DEVICE RECEIVED WITH CRITICAL PUMP ERROR AND PUMP ERROR 35 DUE TO MOISTURE DAMAGE TO FORCE SENSOR. DEVICE RECEIVED WITH CORRODED ELECTRONIC ASSEMBLIES AND CORRODED MOTOR HOME SWITCH. UNABLE TO PERFORM DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT AND SELF TEST DUE TO CRITICAL PUMP ERROR ALARM NOTED.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP RECEIVED AN ALARM WHICH INDICATED THAT A BROKEN FORCE SENSOR WAS DETECTED DURING SEATING OR REGULAR DELIVERY. CUSTOMER WAS IN WATER WHEN INSULIN PUMP START BEEPING. CUSTOMER NOTICED THAT THE WATER WENT INTO THE BATTERY COMPARTMENT. SCREEN DID NOT APPEARED ON THE DISPLAY. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
984316 PUMP MMT-1780KR 670G MG MEDI RECERT PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1780KR HG4DHXT 000000763000365912

Patients

Seq Age Sex Outcome Treatment
1 69 YR