FDA Adverse Event Malfunction Summary report: N

SWAN-GANZ CONTINUOUS CARDIAC OUTPUT CATHETER

MDR report key: 1051932 · Received May 20, 2008

Report

Report Number
6000002-2008-07265
Event Type
Malfunction
Date Received
May 20, 2008
Date of Event
March 13, 2008
Report Date
March 17, 2008
Manufacturer
EDWARDS LIFESCIENCES, PR
Product Code
DYG
PMA / PMN Number
K924452
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED AND EVALUATED. CUSTOMER REPORT WAS CONFIRMED. BALLOON LUMEN LEAKAGE OBSERVED THROUGH A SLIT OF THE CATHETER BODY, 0.06" IN LENGTH, AT THE 12.5 CM AREA (DISTAL OF THE THERMAL FILAMENT). NO VISIBLE DAMAGE TO BALLOON LATEX. A CAPA IS IN PROCESS FOR SLIT IN CATHETER BODY RELATED TO BALLOON INFLATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BALLOON DID NOT INFLATE AND THE BALLOON COULD NOT MAINTAIN INFLATION. NO PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SWAN-GANZ CONTINUOUS CARDIAC OUTPUT CATHETER CONTINUOUS CARDIAC OUTPUT CATHETER DYG EDWARDS LIFESCIENCES, PR 744HF75 58430878

Patients

Seq Age Sex Outcome Treatment
1 UNK Other