FDA Adverse Event
Malfunction
Summary report: N
SWAN-GANZ CONTINUOUS CARDIAC OUTPUT CATHETER
MDR report key: 1051932
·
Received May 20, 2008
Report
- Report Number
- 6000002-2008-07265
- Event Type
- Malfunction
- Date Received
- May 20, 2008
- Date of Event
- March 13, 2008
- Report Date
- March 17, 2008
- Manufacturer
- EDWARDS LIFESCIENCES, PR
- Product Code
- DYG
- PMA / PMN Number
- K924452
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED AND EVALUATED. CUSTOMER REPORT WAS CONFIRMED. BALLOON LUMEN LEAKAGE OBSERVED THROUGH A SLIT OF THE CATHETER BODY, 0.06" IN LENGTH, AT THE 12.5 CM AREA (DISTAL OF THE THERMAL FILAMENT). NO VISIBLE DAMAGE TO BALLOON LATEX. A CAPA IS IN PROCESS FOR SLIT IN CATHETER BODY RELATED TO BALLOON INFLATION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE BALLOON DID NOT INFLATE AND THE BALLOON COULD NOT MAINTAIN INFLATION. NO PATIENT COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SWAN-GANZ CONTINUOUS CARDIAC OUTPUT CATHETER | CONTINUOUS CARDIAC OUTPUT CATHETER | DYG | EDWARDS LIFESCIENCES, PR | 744HF75 | 58430878 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Other |