FDA Adverse Event Malfunction Summary report: N

MED-EL COMBI 40+ COCHLEAR IMPLANT SYSTEM

MDR report key: 1051886 · Received May 13, 2008

Report

Report Number
9710014-2008-00164
Event Type
Malfunction
Date Received
May 13, 2008
Date of Event
April 29, 2008
Report Date
May 13, 2008
Manufacturer
MED-EL ELEKTROMEDIZINISCHE GERAETE, GMBH
Product Code
MCM
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN EXPLANTED. IF IT SHOULD BE EXPLANTED, IT IS TO BE RETURNED TO THE MFR FOR EVAL. WHEN AVAILABLE, A DEVICE FAILURE ANALYSIS WILL BE SUBMITTED AS A F/U REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING, IT WAS SEEN THAT 9 OF THE 12 ELECTRODE CHANNELS HAD HIGH IMPEDANCES. THE PT'S HEARING ABILITY HAS BECOME LESS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MED-EL COMBI 40+ COCHLEAR IMPLANT SYSTEM COCHLEAR IMPLANT MCM MED-EL ELEKTROMEDIZINISCHE GERAETE, GMBH C40+

Patients

Seq Age Sex Outcome Treatment
1 10 YR