CELLSEARCH CIRCULATING TUMOR CELL KIT
Report
- Report Number
- 3004582358-2008-00002
- Event Type
- Malfunction
- Date Received
- May 19, 2008
- Date of Event
- April 18, 2008
- Report Date
- May 16, 2008
- Manufacturer
- VERIDEX, LLC
- Product Code
- NQI
- PMA / PMN Number
- K073338
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
TWO CD LOG FILES OF THE RESULT WERE RETURNED BY THE CUSTOMER TO VERIDEX FOR TECHNICAL REVIEW. ON-SITE VISIT BY VERIDEX REPRESENTATIVE WAS CONDUCTED. NO DEFINITIVE ROOT CAUSE WAS DETERMINED FOR REPORTED DISCREPANT RESULTS (79 CTC AND 4 CTC). THE SYSTEM LOGS PROVIDED BY THE CUSTOMER WERE REVIEWED BY A VERIDEX SPECIALIST AND CONFIRMED THE CTCS OF 79 AND 4 RESPECTIVELY AS IDENTIFIED BY THE CUSTOMER. DURING ON SITE VISIT CONDUCTED BY VERIDEX, SAMPLE PREPARATION AND ANALYSIS WAS OBSERVED. THE OBSERVATION INDICATED THAT THE PROCEDURE BEING USED BY THE CUSTOMER WAS NOT IN ACCORDANCE WITH THE USER TRAINING WITH RESPECT TO SAMPLE SEPARATION.
CUSTOMER REPORTED TWO SAMPLES DRAWN FROM THE SAME PT ON (B) (6) 2008 (AROUND 11AM PST) GENERATED RESULTS OF 79 CIRCULATING TUMOR CELLS (CTC) AND 4 CTC. THE FIRST TEST WAS CONDUCTED THE EVENING OF (B) (6) 2008 USING CTC KIT LOT 0087. THE SECOND TEST WAS CONDUCTED ON (B) (6) 2008 ALSO USING CTC KIT LOT 0087. A CONTROL WAS INCLUDED WITH BOTH TESTS. BLOOD WAS COLLECTED IN THE CELLSAVE TUBE AND STORED AT ROOM TEMPERATURE, (CELLSAVE TUBE LOT NUMBER 725806). NO AGGREGATION WAS OBSERVED. NO OTHER SAMPLES WERE TESTED ON EITHER DAY. THE WASH BUFFER WAS STORED WITH THE KIT IN THE FRIDGE AFTER EACH USE. CONTROLS PASSED. CUSTOMER STATES THAT DAILY CLEANING IS PERFORMED ON THE AUTOPREP. CUSTOMER INDICATED THAT THE 4 CTC CELL COUNT WAS REPORTED. INSTRUMENTS USED WERE AUTOPREP (B) (4) AND CELLTRACKS ANALYZER (B) (4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CELLSEARCH CIRCULATING TUMOR CELL KIT | CELLSEARCH CTC KIT | NQI | VERIDEX, LLC | NA | 0087 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NOT AVAILABLE |