FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 10516667 · Received September 10, 2020

Report

Report Number
1645337-2020-11338
Event Type
Injury
Date Received
September 10, 2020
Date of Event
July 28, 2020
Report Date
August 20, 2020
Manufacturer
MENTOR TEXAS
Product Code
FTR
UDI-DI
00081317000709
PMA / PMN Number
P030053
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON (B)(6) 2020, DURING VISUAL EVALUATION OF THE DEVICE NO APPARENT DAMAGE OR ANOMALIES WERE OBSERVED. LEAK TESTING WAS PERFORMED, ACCORDING WITH MENTOR PROCEDURE, AND IT REVEALED A TEAR WITHIN THE SILTEX CRACKING MEASURING APPROXIMATELY 0.2 CM IN THE POSTERIOR VIEW. SILTEX CRACKING IS DEFINED AS A MATERIAL SEPARATION OCCURRING IN THE SILTEX LAYER AND CAN EVENTUALLY PROGRESS THROUGH THE SHELL OF SILTEX DEVICES. SILTEX CRACKING IS ULTIMATELY A RESULT OF HIGH STRESSES. A SECOND PRODUCT WAS RECEIVED. NO ADVERSE EVENTS WERE REPORTED FOR THIS CONCOMITANT (CONTRALATERAL) DEVICE, THEREFORE NO FURTHER ANALYSIS IS REQUIRED. AS PART OF OUR QUALITY PROCESS, THE MANUFACTURING RECORDS OF THIS LOT-SERIAL NUMBER WERE REVIEWED AND THE MANUFACTURING STANDARDS WERE MET PRIOR TO THE RELEASE OF THIS LOT. EACH DEVICE IS VISUALLY INSPECTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. RUPTURE COMPLAINT INFORMATION IS CONSISTENTLY ANALYZED AND MONITORED BY MENTOR QUALITY ASSURANCE TO DETERMINE WHEN FURTHER ACTION IS NECESSARY. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON (B)(6) 2020, A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED FOR THE FINISHED DEVICE NUMBER 150428, AND NO NON-CONFORMANCE'S RELATED TO THE REPORTED COMPLAINT WERE IDENTIFIED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. A MANUFACTURING RECORD EVALUATION IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: RUPTURE. MEDICAL DEVICE REMOVAL. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6)-YEAR-OLD FEMALE PATIENT UNDERWENT A BREAST AUGMENTATION REVISION WITH A MENTOR MEMORYGEL BREAST IMPLANT 375CC AND EXPERIENCED RUPTURE ON THE RIGHT BREAST IMPLANT. AS A RESULT, THE PATIENT HAD UNDERGONE REMOVAL AND REPLACEMENT WITH MENTOR MEMORYGEL BREAST IMPLANT 375CC ON (B)(6) 2020.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
979637 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 3543757 150428 00081317000709

Patients

Seq Age Sex Outcome Treatment
1 67 YR Required Intervention