HAKIM PROGRAMMABLE VALVE
Report
- Report Number
- 1226348-2008-00129
- Event Type
- Injury
- Date Received
- May 21, 2008
- Date of Event
- April 18, 2008
- Manufacturer
- CODMAN & SHURTLEFF, INC./MEDOS S.A.
- Product Code
- JXG
- PMA / PMN Number
- K980778
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
CODMAN HAS REQUESTED RETURN OF THE VALVE FOR EVALUATION. HISTORICALLY, FOR COMPLAINTS OF THIS NATURE, EXAMINATIONS OF THE VALVES HAVE REVEALED THE ACCUMULATIONS OF BIOLOGICAL DEBRIS WITHIN THE VALVES, WHICH HAVE RESULTED IN BLOCKAGES WITHIN THE DEVICES. REVIEWS OF THE DEVICE HISTORY RECORDS HAVE CONFIRMED THAT THE VALVES HAVE MET SPECIFICATIONS WHEN RELEASED TO STOCK. A FOLLOW UP REPORT WILL BE FILED IF THE INVESTIGATION OF THIS DEVICE REVEALS A RESULT DIFFERENT FROM THAT WHICH IS STATED ABOVE OR, IF ANY FURTHER INFO REGARDING THE CAUSE OR MODE OF FAILURE IS MADE AVAILABLE. OTHERWISE, THE COMPLAINT IS CLOSED AT THIS TIME.
AFFILIATE REPORTED THAT FLUID DID NOT FLOW THROUGH THE DEVICE AND IT NEEDED TO BE REPLACED. IT WAS IMPLANTED IN 2007 AND REPLACED IN 2008.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HAKIM PROGRAMMABLE VALVE | SHUNT, CENTRAL NERVOUS SYSTEM & COMPS | JXG | CODMAN & SHURTLEFF, INC./MEDOS S.A. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |