ELECSYS 2010 RACK
Report
- Report Number
- 1823260-2008-04147
- Event Type
- Malfunction
- Date Received
- May 19, 2008
- Date of Event
- April 25, 2008
- Report Date
- May 19, 2008
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- DHA
- PMA / PMN Number
- K961481
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
USER RECEIVED DISCREPANT PT HCG RESULTS FOR MULTIPLE PTS. ONLY THE FOLLOWING FIVE PT EXAMPLES WERE PROVIDED. PT 1: IN 2008, INITIAL RESULT GAVE >10000 MIU/ML, SAMPLE REPEATED THREE TIMES WITH DILUTION GIVING 1158, 1484 AND 68590 MIU/ML. PT 2: IN 2008, INITIAL RESULT GAVE >10000 MIU/ML; SAMPLE REPEATED TWICE WITH DILUTION SAME DAY GAVE 3213 MIU/ML, AND AGAIN IN THE SAME DAY, GAVE 14250 MIU/ML. PT 3, IN 2008, INITIAL RESULT GAVE >10000 MIU/ML, SAMPLE REPEATED TWICE WITH DILUTION, SAME DAY GAVE 7589 MIU/ML, AND AGAIN ABOUT 4 DAYS LATER, GAVE 20330 MIU/ML. PT 4, IN 2008, INITIAL RESULT GAVE 2058 MIU/ML, REPEAT GAVE 455.2 MIU/ML. ERRONEOUS RESULTS WERE REPORTED. PTS NOT ADVERSELY AFFECTED. THE FIELD SERVICE REP DETERMINED THE CAUSE OF THE DISCREPANT RESULTS TO BE DUE TO A PROBLEM WITH THE Z-MOTOR, AND S/R RIBBON CABLE, AND REPLACED Z-MOTOR AND S/R RIBBON CABLE.
USER RECEIVED DISCREPANT PT HCG RESULTS FOR MULTIPLE PTS. ONLY THE FOLLOWING FIVE PT EXAMPLES WERE PROVIDED. PT 5: IN 2008, INITIAL RESULT GAVE >10000 MIU/ML, SAMPLE REPEATED TWICE WITH DILUTION, SAME DAY GAVE 8800 MIU/ML AND NEXT DAY GAVE 26760 MIU/ML. ERRONEOUS RESULTS WERE REPORTED. PTS NOT ADVERSELY AFFECTED. THE FIELD SERVICE REP DETERMINED THE CAUSE OF THE DISCREPANT RESULTS TO BE DUE TO A PROBLEM WITH THE Z-MOTOR AND S/R RIBBON CABLE, AND REPLACED Z-MOTOR AND S/R RIBBON CABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ELECSYS 2010 RACK | IMMUNOCHEMISTRY ANALYZER - JJE | DHA | ROCHE DIAGNOSTICS | 2010 RACK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | |||
| 2 | 24 YR |