CELL-DYN SAPPHIRE ANALYZER
Report
- Report Number
- 2919069-2008-00547
- Event Type
- Malfunction
- Date Received
- May 21, 2008
- Date of Event
- April 28, 2008
- Report Date
- May 6, 2008
- Manufacturer
- ABBOTT DIAGNOSTICS DIVISION/CELLDYN
- Product Code
- GKZ
- PMA / PMN Number
- K051215
- Removal / Correction Number
- Z-1165-2008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
(B)(4). CUSTOMERS REPORTED THAT THE HEMOGLOBIN BACKGROUND COUNTS WERE OUT OF THE SPECIFICATION HIGH AFTER INSTALLING AND RUNNING THE CELL-DYN SAPPHIRE HEMOGLOBIN REAGENT SYRINGE LIST NUMBER 08H49-02 ON THE CELL-DYN SAPPHIRE INSTRUMENT. THE SPECIFICATION FOR THE HEMOGLOBIN BACKGROUND IS LESS THAN OR EQUAL TO 0.10 GRAMS PER DECILITERS PER THE CELL-DYN SAPPHIRE OPERATORS MANUAL. BACKGROUND COUNTS MUST BE WITHIN SPECIFICATION PRIOR TO REPORTING OF PATIENT SAMPLES. INVESTIGATION OF THIS ISSUE FOUND THAT THE ELEVATED HGB BACKGROUND COUNT IS DUE TO EXCESS LUBRICANT USED BY THE VENDOR DURING THEIR MANUFACTURING PROCESS OF THE CELL-DYN SAPPHIRE HEMOGLOBIN REAGENT SYRINGE THAT IS PRESENT ON THE TIP OF THE PLUNGER, LUBRICANT SHOULD ONLY BE APPLIED TO THE SIDE OF THE PLUNGER. THE CELL-DYN SAPPHIRE HEMOGLOBIN REAGENT SYRINGE LIST NUMBER 08H49-02 IS DISTRIBUTED FOR USE ON THE CELL-DYN SAPPHIRE INSTRUMENT ONLY. THIS ISSUE HAS NO IMPACT ON OTHER PLATFORMS OR INSTRUMENTS. THE FOLLOWING CORRECTIVE ACTIONS WERE PUT IN PLACE IN ORDER TO ADDRESS THE ISSUE A PRODUCT RECALL LETTER FA11JAN2008 WAS ISSUED WITH ITS ASSOCIATED FIELD ACTION AND DISTRIBUTED TO AFFECTED CUSTOMERS. AFFECTED CELL-DYN SAPPHIRE HEMOGLOBIN REAGENT SYRINGES WERE IDENTIFIED TO HAVE PACKAGED DATES OF (B)(6) 2007 THROUGH (B)(6) 2007 FOR FA11JAN2008. THE VENDOR, (B)(4), HAS IMPLEMENTED AND UPDATED THEIR PROCEDURES IN REGARDS TO THE APPLICATION OF THE LUBRICANT AND CLEANING OF THE CELL-DYN SAPPHIRE HEMOGLOBIN REAGENT SYRINGE AS PART OF THEIR MANUFACTURING PROCESS TO ADDRESS THE EXCESS LUBRICANT ISSUE. ABBOTT (B)(4) HAS IMPLEMENTED AN INCOMING QUALITY INSPECTION REQUIREMENT FOR THE CELL-DYN SAPPHIRE HEMOGLOBIN REAGENT SYRINGE PRIOR TO TESTING OR USE ON THE MANUFACTURING LINE. THIS IS A FINAL REPORT.
THE CUSTOMER IS QUESTIONING THE HIGH HEMOGLOBIN BACKGROUNDS GENERATED ON THE CELL-DYN SAPPHIRE ANALYZER USING THE CELL-DYN SAPPHIRE HEMOGLOBIN SYRINGE, LIST # 8H49-02. THE CUSTOMER PERFORMED AN AUTO CLEAN TWICE, CLEANED THE SYRINGE, FLUSHED THE HEMOGLOBIN SYRINGE PATHWAY AND REPLACED THE HEMOGLOBIN REAGENT WITH NO RESOLUTION. THE ISSUE WAS RESOLVED BY REPLACING THE HEMOGLOBIN SYRINGE. THERE WAS NO IMPACT TO PT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CELL-DYN SAPPHIRE ANALYZER | AUTOMATED HEMATOLOGY ANALYZER | GKZ | ABBOTT DIAGNOSTICS DIVISION/CELLDYN | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI | LIST # 8H49-02| CELL-DYN SAPPHIRE HEMOGLOBIN SYRINGE| CELL-DYN SAPPHIRE HEMOGLOBIN SYRINGE, LN 8H49-02 |