FDA Adverse Event Malfunction Summary report: N

CASE-15ML SYRINGE WITH HANDIFIL

MDR report key: 1051365 · Received May 16, 2008

Report

Report Number
9610849-2008-00018
Event Type
Malfunction
Date Received
May 16, 2008
Report Date
August 2, 2005
Manufacturer
COVIDIEN
Product Code
FMF
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: NO PHYSICAL SAMPLE WAS RECEIVED FOR EVALUATION. THERE WAS A COMPLETE REVIEW OF THE OPERATIONAL CONDITIONS OF THE L F MANUFACTURING LINE FROM MARCH 2005 TO MAY 2005 WERE BOTH EQUIPMENT AND PERSONNEL CLEANING AND GOWNING PRACTICES, IDENTIFYING THE POTENTIAL SOURCES OF FOREIGN OBJECTS WHICH WERE ADDRESSED AS LISTED BELOW. CONCLUSIONS/CORRECTIVE ACTIONS: THE NEXT ACTIONS WERE TAKEN TO PREVENT THE FOREIGN OBJECTS IN PRODUCT OR ITS PACKAGE. REINFORCE THE 6S PROGRAM IN MANUFACTURING AREA, WHICH CONSISTS ON DAILY CLEANING ROUTINES BY OUR MANUFACTURING PERSONNEL DURING SHIFT TRANSITIONS. MATERIAL HANDLING PRACTICES WERE CHANGED TO PREVENT THE ADHESION OF FOREIGN PARTICLES TO COMPONENTS SUCH AS THE ROLL STOCK WHICH IS THE RAW MATERIAL FOR THE SYRINGE TRAY. STICKY MATS WERE INSTALLED IN THE ENTRANCE DOOR OF THE MANUFACTURING AREA TO COLLECT THE DUST FORM THE PERSONNEL SHOES BEFORE THEY GET INTO THE ASSEMBLY AREA. GOWNING PRACTICES WERE REVIEWED WITH ALL LF ASSEMBLY LINE PERSONNEL TO REINFORCE ITS CORRECT APPLICATION WHICH INCLUDE THE ADEQUATE USE AND REPLACEMENT OF HEAD COVERS.

Description of Event or Problem · 1

REPORTER REPORTS: WHEN WE HAD OUR INSPECTION FOR OUR STOCK, A CHIP OF PAPER LIKE OBJECT WAS FOUND INSIDE THE SYRINGE BARREL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CASE-15ML SYRINGE WITH HANDIFIL DISPOSABLE SYRINGE, EMPTY FMF COVIDIEN 600269 5101132

Patients

Seq Age Sex Outcome Treatment
1 UNK