FDA Adverse Event Other Summary report: N

ESPIRIT VENTILATOR

MDR report key: 1051336 · Received May 21, 2008

Report

Report Number
2031642-2008-00114
Event Type
Other
Date Received
May 21, 2008
Date of Event
April 23, 2008
Report Date
April 23, 2008
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
CBK
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE VENTILATOR BEGAN ALARMING 'LOW BATTERY' AND THEN SHUT DOWN WHILE IN USE ON A PATIENT. THE CUSTOMER REPORTED THERE WAS NO PATIENT HARM. THE MANUFACTURER'S SERVICE TECHNICIAN REPORTED THE VENTILATOR'S DIAGNOSTIC LOG SHOWED THE DEVICE HAD BEEN RUNNING ON BACKUP BATTERY POWER WHICH BECAME DEPLETED DURING EXTENDED USE. THE BACKUP BATTERY FUNCTIONED UNTIL IT DEPLETED CAUSING THE VENTILATOR TO SHUT DOWN AND ALARM AS SPECIFIED. APPLICABLE TESTING WAS COMPLETED AND ALL TESTS PASSED PER SPECIFICATIONS. THERE WAS NO MALFUNCTION OF THE DEVICE OR BACKUP BATTERY. THIS EVENT IS BEING REPORTED AS A USER ERROR. PER THE ESPIRIT OPERATORS MANUAL, WHEN THE VENTILATOR IS POWERED BY BACKUP BATTERY IT WILL GENERATE A NON-RESETTABLE, NONSILENCEABLE ALARM EVERY 60 SECONDS. DURING THIS STATE A BATTERY IN USE YELLOW LED IS LIT. OPERATION WILL CONTINUE UNTIL THE BATTERY HAS 5 MINUTES OF OPERATION LEFT. A RED BATTERY LOW LED WILL FLASH AND A NON-RESETTABLE HIGH PRIORITY ALARM WILL SOUND. WHEN THE BATTERY LOW INDICATOR IS FLASHING RED, OPERATION OF THE VENTILATOR FROM THE BATTERY POWER SHOULD BE DISCONTINUED. THE BACKUP BATTERY WILL CHARGE WHEN THE ESPIRIT IS PLUGGED INTO A VIABLE AC SUPPLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ESPIRIT VENTILATOR VENTILATOR, CONTINUOUS CBK RESPIRONICS CALIFORNIA, INC. V1000 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK