FDA Adverse Event Malfunction Summary report: N

VOYAGER OTW CORONARY DILATATION CATHETER

MDR report key: 1051192 · Received May 22, 2008

Report

Report Number
2024168-2008-00415
Event Type
Malfunction
Date Received
May 22, 2008
Date of Event
April 22, 2008
Report Date
April 24, 2008
Manufacturer
ABBOTT VASCULAR-CARDIAC THERAPIES
Product Code
LOX
PMA / PMN Number
P810046
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESULTS AND CONCLUSION SUMMATION: PRODUCT PERFORMANCE ENGINEERING DETERMINED THAT THE FACTORS THAT CAN CONTRIBUTED TO BALLOON RUPTURE INCLUDE, BUT ARE NOT LIMITED TO, BALLOON DAMAGE DURING MANUFACTURING, MATERIALS, INTERACTIONS WITH OTHER DEVICES, PATIENT ANATOMY, LESION CALCIFICATION, LESION TORTUOSITY, EXCESSIVE APPLIED PRESSURE, OR INSUFFICIENT PREPARATION PRIOR TO USE. IT WAS REPORTED THAT THE LESION TREATED IN THE PROCEDURE WAS MILDLY CALCIFIED, WHICH COULD HAVE CONTRIBUTED TO DAMAGING THE SURFACE OF THE BALLOON SUCH THAT UPON INFLATION, THE BALLOON RUPTURED; HOWEVER, WITHOUT A RETURNED DEVICE FOR ANALYSIS, A DEFINITE ROOT CAUSE FOR THE BALLOON RUPTURE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

REPORTING STATUS: MALFUNCTION. REPORTING RATIONALE: BALLOON RUPTURE IS LIKELY TO CAUSE OR CONTRIBUTE TO PATIENT INJURY. DEVICE ISSUE: BALLOON RUPTURE. IT WAS REPORTED THAT THE VOYAGER OTW WAS CROSSED TO THE LESION; HOWEVER, IT RUPTURED AT 3 ATM, WHEN IT WAS INFLATED FOR THE FIRST TIME. THE LESION WAS DILATED WITH A 1.5 X 15 MM VOYAGER RX BALLOON CATHETER AND THEN WITH A 3.0 X 15 MM NC MERCURY. THE PROCEDURE WAS COMPLETED AFTER DEPLOYING A 3.0 X 18 MM VISION STENT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VOYAGER OTW CORONARY DILATATION CATHETER LOX ABBOTT VASCULAR-CARDIAC THERAPIES NA 7012951

Patients

Seq Age Sex Outcome Treatment
1 UNK STENT: VISION 3.0-18| GUIDE CATHETER: MACH1 JR4| DILATATION CATHETER: VOYAGER RX 1.5 X 15| GUIDE WIRE: PILOT 50