ALARIS PC UNIT
Report
- Report Number
- 2016493-2020-05168
- Event Type
- Malfunction
- Date Received
- September 9, 2020
- Report Date
- August 13, 2020
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- UDI-DI
- 10885403801518
- PMA / PMN Number
- K091308
- Removal / Correction Number
- Z-2909-2020
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
SUPPLEMENTAL CREATED TO REVISE H6 EVALUATION AND CONCLUSION CODES TO REFLECT NO FAULT FOUND-ADDITIONAL INFORMATION ADDED TO: H6 H3 OTHER TEXT : FI ADDED TO PREVIOUS MDR.
THE CUSTOMER REPORTED THAT THE FRONT CASE WITH KEYPAD IS BEING REPLACED WITH RECALL REPLACEMENT PARTS. THE RETURNED FRONT CASE W/KEYPAD ASSEMBLIES WERE CONFIRMED NEW AND ARE LISTED IN THE MEDICAL DEVICE RECALL LETTER SENT TO THE CUSTOMER ON 04 AUG 2020. EXTERNAL INSPECTION WAS PERFORMED ON THE 7 PRINTEC (P/N TC10013702) KEYPADS DATE CODE 18/01 (JAN 2018). THE KEYPADS WERE OBSERVED TO BE IN NEW UNUSED CONDITION WITH NO IRREGULARITIES OBSERVED. TEST RESULTS IDENTIFIED THE KEYPADS TESTED GOOD WITH NO FAULTS IDENTIFIED. TEST RESULTS: MAINTENANCE KEYPAD TEST: THE FRONT CASE KEYPAD (QTY 7 P/N TC10013702) WERE CONNECTED TO THE CAD PCU TEST FIXTURE #10013973 (EQ 111582) FOR FUNCTIONAL TESTING. THE KEYPAD¿S AC INDICATOR LIGHT ILLUMINATED AS EXPECTED AND ALL KEYS FUNCTIONED AS INTENDED. TEST METHOD INFORMATION: N/A. N/A. THE RETURNED FRONT CASE W/KEYPADS ARE CONFIRMED NEW AND UNUSED. THE ASSEMBLIES WERE TESTED WITH NO IRREGULARITIES FOUND. THE ASSEMBLIES ARE BEING RETURNED PER RECALL FOR REPLACEMENT PARTS. REVIEW OF THE PCU MODULE (B)(6) SERVICE HISTORY RECORD COULD NOT BE PERFORMED AS S/N PROVIDED IS NOT A VIAL S/N. REVIEW OF THE PRODUCTION FAILURE RECORD RECORD COULD NOT BE PERFORMED AS S/N PROVIDED IS NOT A VALID S/N. H3 OTHER TEXT: ONLY KEYPADS WERE RECEIVED FOR INVESTIGATION.
IT WAS REPORTED THE FRONT CASE WITH KEYPAD IS BEING REPLACED. THERE WAS NO PATIENT INVOLVEMENT.
IT WAS REPORTED THE FRONT CASE WITH KEYPAD IS BEING REPLACED. THERE WAS NO PATIENT INVOLVEMENT.
NO DEVICE WILL BE RETURNED PER CUSTOMER. THE CUSTOMER COMPLAINT COULD NOT BE CONFIRMED BECAUSE THE DEVICE WAS NOT RETURNED FOR FAILURE INVESTIGATION. THE ROOT CAUSE OF THIS FAILURE WAS NOT IDENTIFIED.
IT WAS REPORTED THE FRONT CASE WITH KEYPAD IS BEING REPLACED. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 974209 | ALARIS PC UNIT | PUMP, INFUSION | FRN | CAREFUSION SD | 8015 | 10885403801518 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |