UNKNOWN DEPUY TIBIAL INSERT
Report
- Report Number
- 1818910-2008-02106
- Event Type
- Injury
- Date Received
- May 21, 2008
- Date of Event
- May 17, 2008
- Report Date
- May 17, 2008
- Manufacturer
- DEPUY ORTHOPAEDICS, INC.
- Product Code
- HSH
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
EVAL WAS NOT POSSIBLE, AS THE PROD WAS NOT RETURNED. PROD INFO REQUIRED TO REVIEW THE DEVICE HISTORY RECORDS WAS NOT PROVIDED. THE INVESTIGATION WAS LIMITED TO THE INFO PROVIDED. THE INVESTIGATION COULD NOT VERIFY OR DRAW ANY CONCLUSIONS ABOUT REPORTED POLYETHYLENE WEAR; HOWEVER, PROVIDED INFO STATES INTERNAL ROTATION OF THE TIBIAL TRAY COMPONENT WAS A CONTRIBUTING FACTOR. THE LENGTH OF TIME IMPLANTED (12 1/2 YRS) IN COMBINATION WITH THE REPORTED PT ACTIVITY LEVEL COULD ALSO BE CONTRIBUTING FACTORS. BASED ON THE INVESTIGATION FINDINGS, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PROD AND/OR ADD'L INFO BE REC'D, THE INVESTIGATION WILL BE RE-OPENED.
PT REVISED TO ADDRESS POLYETHYLENE WEAR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNKNOWN DEPUY TIBIAL INSERT | TOTAL KNEE REPLACEMENT | HSH | DEPUY ORTHOPAEDICS, INC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Required Intervention |