FDA Adverse Event
Injury
Summary report: N
RIATA ST OPTIM PASSIVE FIXATION LEAD
MDR report key: 1050918
·
Received May 27, 2008
Report
- Report Number
- 2017865-2008-01857
- Event Type
- Injury
- Date Received
- May 27, 2008
- Date of Event
- March 18, 2008
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWS
- PMA / PMN Number
- PMA00001
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT AT FOLLOW-UP, THE PATIENT COMPLAINED OF TWITCHING. MEASURED DATA SHOWED UNDERSENSING AND A DROP IN IMPEDANCE. THRESHOLD COULD NOT BE CAPTURED AND THE TWITCHING WAS INDUCED DURING PACING. A PERFORATION WAS CONFIRMED BY X-RAY, ULTRASONOGRAPHY AND CT. AT EXPLANT, AN INFECTION WAS ALSO NOTED. THE LEAD WAS DISCARDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RIATA ST OPTIM PASSIVE FIXATION LEAD | NO BNCN FOUND FOR THIS DEVICE | LWS | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7070/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |