FDA Adverse Event
Injury
Summary report: N
RIATA PASSIVE FIXATION LEAD
MDR report key: 1050845
·
Received May 27, 2008
Report
- Report Number
- 2017865-2008-01792
- Event Type
- Injury
- Date Received
- May 27, 2008
- Date of Event
- March 11, 2008
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- LWS
- PMA / PMN Number
- PMA00001
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY DUE TO NOISE ON THE VENTRICULAR CHANNEL. PACING LEAD IMPEDANCE WAS FOUND TO BE HIGH AND LOSS OF CAPTURE WAS NOTED. A CONDUCTOR BREAK WAS OBSERVED ON THE LEAD DURING EXPLANT. THE LEAD WILL NOT BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RIATA PASSIVE FIXATION LEAD | DEFIBRILLATION LEAD | LWS | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1570/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |