FDA Adverse Event Injury Summary report: N

RIATA ST OPTIM ACTIVE FIXATION

MDR report key: 1050798 · Received May 27, 2008

Report

Report Number
2017865-2008-01750
Event Type
Injury
Date Received
May 27, 2008
Date of Event
February 6, 2008
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
LWS
PMA / PMN Number
PMA00001
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED HV INAPPROPRIATE THERAPY. THE DEVICE WAS DOUBLE COUNTING THE P-WAVE ON THE VENTRICULAR CHANNEL. LEAD DISLODGEMENT WAS SUSPECTED. THE X-RAYS CONFIRMED THE RV COIL ALL THE WAYUP INTO THE SVC. THE LEAD WAS REPOSITIONED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RIATA ST OPTIM ACTIVE FIXATION NO BNCN FOUND FOR THIS DEVICE LWS ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7020/60 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention