FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES

MDR report key: 10505315 · Received September 8, 2020

Report

Report Number
1917413-2020-00790
Event Type
Malfunction
Date Received
September 8, 2020
Date of Event
August 18, 2020
Report Date
September 9, 2020
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
JKA
UDI-DI
50382903678865
PMA / PMN Number
K945952
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FROM THE CUSTOMER FOR INVESTIGATION. THEREFORE, 10 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY DRAW TESTING AND NO ISSUES WERE OBSERVED RELATING TO UNDERFILL AS ALL SAMPLES MET SPECIFICATIONS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE TUBES ARE UNDER FILLING DURING USE WITH A BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER CALLED IN AND STATED THAT THE VACUTAINER IS NOT FILLING ALL THE WAY UP TO THE GREEN LINE. THE ITEM NUMBER IS 367886, LOT#: 9220481. SHE STATED THEY NOTICED THIS ISSUE A COUPLE OF MONTHS AGO AND SHE DOES NOT KNOW IF ANY ERRONEOUS RESULTS HAPPENED BECAUSE OF THIS.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INITIAL REPORTER STATE: ADDRESS INFORMATION WAS NOT ABLE TO BE OBTAINED. (B)(4) WAS USED AS A PLACE HOLDER BASED ON THE USER AREA CODE. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE TUBES ARE UNDER FILLING DURING USE WITH A BD VACUTAINER® LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER CALLED IN AND STATED THAT THE VACUTAINER IS NOT FILLING ALL THE WAY UP TO THE GREEN LINE. THE ITEM NUMBER IS 367886, LOT#: 9220481. SHE STATED THEY NOTICED THIS ISSUE A COUPLE OF MONTHS AGO AND SHE DOES NOT KNOW IF ANY ERRONEOUS RESULTS HAPPENED BECAUSE OF THIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
970223 BD VACUTAINER LITHIUM HEPARINN (LH) 95 USP UNITS BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. (BROKEN BOW) 367886 9220481 50382903678865

Patients

Seq Age Sex Outcome Treatment
1 Other