FDA Adverse Event Malfunction Summary report: N

G3 PLUS AUTO, AED, AHA 2010, BRITISH

MDR report key: 10504651 · Received September 8, 2020

Report

Report Number
2112020-2020-00688
Event Type
Malfunction
Date Received
September 8, 2020
Report Date
August 18, 2020
Manufacturer
CARDIAC SCIENCE CORPORATION
Product Code
MKJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GB
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION FOR EVALUATION. THE CUSTOMER'S REPORT WAS OBSERVED DURING REVIEW OF THE DEVICE DATA LOGS. HOWEVER, THE DEVICE WAS PUT THROUGH EXTENSIVE TESTING INCLUDING FULL FUNCTIONAL TESTING AND DWM TESTING WITHOUT DUPLICATING THE REPORT. THE REPORT WAS CLEARED FROM THE LOGS. AN INTERNAL INSPECTION FOUND NO DISCREPANCIES. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS OF REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A SUPPLEMENTAL REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DETECTED HIGH IMPEDANCE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
969913 G3 PLUS AUTO, AED, AHA 2010, BRITISH AUTOMATED EXTERNAL DEFIBRILLATOR MKJ CARDIAC SCIENCE CORPORATION 9390A-1002 NA

Patients

Seq Age Sex Outcome Treatment
1