FDA Adverse Event Malfunction Summary report: N

BD 10ML SYRINGE LUER-LOK TIP

MDR report key: 10502655 · Received September 8, 2020

Report

Report Number
1213809-2020-00605
Event Type
Malfunction
Date Received
September 8, 2020
Date of Event
August 17, 2020
Report Date
September 24, 2020
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903009122
PMA / PMN Number
K980987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10. DEVICE AVAILABLE FOR EVALUATION?: YES. D.10. RETURNED TO MANUFACTURER ON:(B)(6) 2020. H.6. INVESTIGATION: ONE 10ML SYRINGE IN AN OPENED BLISTER PACK FROM BATCH 9233256 WAS RECEIVED AND EVALUATED. IT WAS OBSERVED THERE WAS SOME BROWN PARTICULATE OUTSIDE THE FLUID PATH, ATTACHED TO THE BOTTOM OF THE PLUNGER ROD. THE PARTICULATE APPEARED TO BE OIL AND WAS LARGE ENOUGH IN SIZE TO BE REJECTABLE PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE FOREIGN MATTER DEFECT IS ASSOCIATED WITH THE MOLDING PROCESS. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF BD 10ML SYRINGE LUER-LOK¿ TIPS EXPERIENCED FOREIGN MATTER CONTAMINATION WHICH WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: BROWN DIRT INSIDE THE SYRINGE.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF BD 10ML SYRINGE LUER-LOK¿ TIPS EXPERIENCED FOREIGN MATTER CONTAMINATION WHICH WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: BROWN DIRT INSIDE THE SYRINGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
970109 BD 10ML SYRINGE LUER-LOK TIP PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 300912 9233256 30382903009122

Patients

Seq Age Sex Outcome Treatment
1 Other