T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY
Report
- Report Number
- 3013756811-2020-96300
- Event Type
- Malfunction
- Date Received
- September 4, 2020
- Date of Event
- August 18, 2020
- Report Date
- September 4, 2020
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00850006613205
- PMA / PMN Number
- DEN180058
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- OTHER
Narratives
PER TANDEM USER GUIDE: ONLY USE U-100 HUMALOG OR U-100 NOVOLOG WITH YOUR PUMP. ONLY U-100 HUMALOG AND NOVOLOG HAVE BEEN TESTED AND FOUND TO BE COMPATIBLE FOR USE IN THE PUMP. USE OF INSULIN WITH LESSER OR GREATER CONCENTRATION CAN RESULT IN UNDER DELIVERY OR OVER DELIVERY OF INSULIN. THIS CAN CAUSE HYPOGLYCEMIA (LOW BG) OR HYPERGLYCEMIA (HIGH BG) EVENTS. NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT OCCLUSION ALARMS OCCURRED. SYSTEM CHECK WAS PERFORMED WITH TANDEM TECHNICAL SUPPORT, AND IT WAS DISCOVERED OCCLUSION WAS CAUSED BY A BLOCKAGE IN THE TUBING. HOWEVER, CUSTOMER REPORTED USING HUMALOG U500 INSULIN. TANDEM TECHNICAL SUPPORT INFORM CUSTOMER THAT HUMALOG U500 IS OFF LABEL PER THE USER GUIDE. CUSTOMER CHANGED PUMP SUPPLIES TO ADDRESS THE ISSUE AND RESUMED INSULIN DELIVERY. CUSTOMER'S BLOOD GLUCOSE WAS 160-281 MG/DL AT TIME OF EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 960289 | T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1002717 | 00850006613205 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR |