CHANGE HEALTHCARE RADIOLOGY SOLUTIONS
Report
- Report Number
- 8022257-2020-00001
- Event Type
- Malfunction
- Date Received
- September 4, 2020
- Date of Event
- July 29, 2020
- Report Date
- September 4, 2020
- Manufacturer
- CHANGE HEALTHCARE CANADA COMPANY
- Product Code
- LLZ
- UDI-DI
- 17540262030020
- PMA / PMN Number
- K140909
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PHYSICIAN
Narratives
CHANGE HEALTHCARE HAS IDENTIFIED A SOFTWARE DEFECT WHICH MAY RESULT IN ONE OR MORE IMAGES/DICOM OBJECTS MISSING FROM A STUDY AFTER IT IS IMPORTED INTO THE PACS. THIS IS CAUSED BY A DISCREPANCY BETWEEN THE IMAGE INDEX AND THE FILES STORED WITHIN THE IMAGE FOLDERS. AS A RESULT OF THIS DISCREPANCY, A USER MAY EXPERIENCE ONE OF THE FOLLOWING SCENARIOS: 1. THE STUDY CANNOT BE OPENED AND AN ERROR MESSAGE IS DISPLAYED, OR 2. THE STUDY CAN BE OPENED AND AN ERROR IS DISPLAYED ON THE VIEWPORT OF THE MISSING IMAGE(S), OR 3. THE STUDY CAN BE OPENED AND NO WARNING MESSAGE IS PRESENTED TO THE USER THAT THERE ARE MISSING IMAGES. THE IMPACTED IMAGE FILES ARE RETAINED IN THE IMAGE FOLDER BUT ARE NOT VISIBLE TO THE USER. WHEN THE STUDY IS ARCHIVED, THESE IMPACTED IMAGE FILES MAY BE DELETED FROM THE PACS PERMANENTLY. CHANGE HEALTHCARE WILL WORK WITH AFFECTED CUSTOMERS TO APPLY A SOFTWARE UPDATE TO PREVENT THE RECURRENCE OF THE PROBLEM.
FOLLOWING AN UPGRADE TO CHRS 14.0 A CUSTOMER SITE, THE CUSTOMER REPORTED ISSUES ON (B)(6) 2020 WHERE SOME NEWLY IMPORTED ULTRASOUND IMAGES WEREN'T IN ORDER AND SOME OF THEM WERE CUT OFF AND BLACK. THERE WAS NO PATIENT HARM REPORTED AS A RESULT OF THIS ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 960586 | CHANGE HEALTHCARE RADIOLOGY SOLUTIONS | PICTURE ARCHIVING AND COMMUNICATIONS SYSTEM | LLZ | CHANGE HEALTHCARE CANADA COMPANY | 14.0 | 17540262030020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |