FDA Adverse Event Injury Summary report: N

COTTON-LEUNG BILIARY STENT

MDR report key: 10493001 · Received September 4, 2020

Report

Report Number
3001845648-2020-00556
Event Type
Injury
Date Received
September 4, 2020
Date of Event
April 6, 2020
Report Date
October 27, 2021
Manufacturer
COOK IRELAND LTD
Product Code
FGE
PMA / PMN Number
K851962
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE EXACT RPN IS UNKNOWN BUT BASED ON THE DEVICE NAME OF COTTON-LEUNG BILIARY STENT, THE TWO POSSIBLE 510K NUMBERS ARE K851962 / K851962. G IMDRF CODE: G04122 - STENT. INVESTIGATION IS STILL PENDING. A FOLLOW UP MDR WILL BE SUBMITTED TO INCLUDE THE INVESTIGATION CONCLUSIONS.

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE COTTON-LEUNG SOFT FLEX BILIARY STENT OF UNKNOWN LOT NUMBER AND RPN INVOLVED IN THIS COMPLAINT WAS NOT AVAILABLE FOR EVALUATION. WITH THE INFORMATION PROVIDED, A DOCUMENT-BASED INVESTIGATION WAS CONDUCTED. THIS FILE WAS CREATED FROM THE JOURNAL ARTICLE. ¿ENDOSCOPIC STENT PLACEMENT ABOVE THE SPHINCTER OF ODDI FOR BILIARY STRICTURES AFTER LIVING DONOR LIVER TRANSPLANTATION". THE BELOW FILES WERE OPENED AS RESULT OF THIS PAPER: (B)(4) (REPORT REFERENCE NUMBER - 3001845648-2020-00555) CREATED TO CAPTURE THE 28 CASES OF THE OFF-LABEL USE OF THE ¿INSIDE STENT¿ (MODIFIED COTTON-LEUNG BILIARY STENT). (B)(4) (REPORT REFERENCE NUMBER - 3001845648-2020-00546) CREATED TO CAPTURE THE 13 STENT MIGRATIONS OF THE OFF-LABEL USE ¿INSIDE STENT¿ (MODIFIED COTTON-LEUNG BILIARY STENT). (B)(4) (REPORT REFERENCE NUMBER - 3001845648-2020-00547) CREATED TO CAPTURE THE 13 CASES STENT OCCLUSION, 3 CASES OF PANCREATITIS AND 14 CASES OF STONE FORMATION OF THE OFF-LABEL USE OF THE ¿INSIDE STENT¿ (MODIFIED COTTON-LEUNG BILIARY STENT). (B)(4) (REPORT REFERENCE NUMBER - 3001845648-2020-00557) CREATED TO CAPTURE 6 CASES OF CHOLANGITIS WITH THE OFF-LABEL USE OF THE ¿INSIDE STENT¿ (MODIFIED COTTON-LEUNG BILIARY STENT). (B)(4) (REPORT REFERENCE NUMBER - 3001845648-2020-00556) CREATED TO CAPTURE 3 CASES PANCREATITIS OF WITH THE USE OF THE ¿OUTSIDE STENT¿ (COTTON-LEUNG BILIARY STENT.) (ON LABEL USE). (B)(4) (REPORT REFERENCE NUMBER - 3001845648-2020-00548) CREATED TO CAPTURE THE 3 CASES OF STONE FORMATION OF THE ¿OUTSIDE STENT¿ (COTTON-LEUNG BILIARY STENT.) (ON LABEL USE). (B)(4) (REPORT REFERENCE NUMBER - 3001845648-2020-00549) CREATED TO CAPTURE THE 4 CASES OF OCCLUSION OF THE ¿OUTSIDE STENT¿ (COTTON-LEUNG BILIARY STENT.) (ON LABEL USE). (B)(4) (REPORT REFERENCE NUMBER - 3001845648-2021-00162) CREATED FOR CASES OF OCCLUSION ASSOCIATED WITH USER ERROR FOR EXCEEDING THE MAXIMUM 3 MONTH INDWELL PERIOD FOR THE ¿OUTSIDE¿ STENTS. (B)(4) (REPORT REFERENCE NUMBER - 3001845648-2021-00161) CREATED FOR THE USER ERROR FOR EXCEEDING THE MAXIMUM 3 MONTH INDWELL PERIOD WITH NO ADVERSE EFFECTS FOR THE ¿OUTSIDE¿ STENT. AS THE RPN AND LOT NUMBER OF THE COMPLAINT DEVICES ARE UNKNOWN, A REVIEW OF THE RELEVANT MANUFACTURING RECORDS CANNOT BE CONDUCTED. HOWEVER, PRIOR TO DISTRIBUTION ALL COTTON-LEUNG SOFT FLEX BILIARY STENT DEVICES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL CHECKS TO ENSURE DEVICE INTEGRITY. THESE INSPECTIONS AND FUNCTIONAL CHECKS ARE OUTLINED IN INTERNAL PROCEDURES IN PLACE AT CIRL. THE CUSTOMER IS NOTIFIED AS PER INSTRUCTIONS FOR USE, (IFU0120-3) WHICH ACCOMPANIES THIS DEVICE, IN THE POTENTIAL COMPLICATIONS SECTION: ¿THOSE ASSOCIATED WITH ERCP INCLUDE BUT ARE NOT LIMITED TO: PANCREATITIS, CHOLANGITIS, ASPIRATION, PERFORATION, HEMORRHAGE, INFECTION, SEPSIS, ALLERGIC REACTION TO CONTRAST OR MEDICATION, HYPOTENSION, RESPIRATORY DEPRESSION OR ARREST, CARDIAC ARRHYTHMIA OR ARREST¿ AND ¿THOSE ASSOCIATED WITH BILIARY STENT PLACEMENT INCLUDE, BUT ARE NOT LIMITED TO: TRAUMA TO THE BILIARY TRACT OR DUODENUM, OBSTRUCTION OF THE PANCREATIC DUCT, STENT MIGRATION.¿ AND IN THE PRECAUTIONS SECTION THE DEVICE ¿SHOULD NOT BE LEFT INDWELLING FOR MORE THAN THREE MONTHS OR AS DIRECTED BY A PHYSICIAN. PERIODIC EVALUATION IS RECOMMENDED.¿ AS PER THE INFORMATION REPORTED BILIARY STENTS WERE PLACED IN 87 SESSIONS: 77 INSIDE SESSIONS AND 10 OUTSIDE SESSIONS. AS BOTH A COMPETITOR STENT AND A COOK DEVICE IS MENTIONED AS USED, FOR THE PURPOSE OF THE WRITE UP A WORST-CASE APPROACH WILL BE TAKEN AND THE COOK DEVICE (COTTON-LEUNG BILIARY STENT) WILL BE ASSUMED AS USED IN EACH CASE. THE ¿OUTSIDE¿ STENT PLACEMENT WAS NOTED AS ON LABEL CONVENTIONAL USE OF THE DEVICE AS THE STENT WAS PLACED ACROSS THE SPHINCTER OF ODDI, WITH THEIR DISTAL END EXPOSED INTO THE DUODENUM. THE ¿INSIDE¿ STENTS WERE NOTED AS PLACED OFF LABEL AND HAD A USER ERROR ASSOCIATED WITH EACH AS THEY WERE MODIFIED WITH A DISTAL FLAP REMOVED AND A NYLON THREAD ATTACHED. THE USER ERROR FOR EXCEEDING THE MAXIMUM IN DWELL PERIOD IS ALSO NOTED HERE AS IT IS REPORTED THAT STENT EXCHANGE/REMOVAL WAS SCHEDULED EVERY 6 TO 12 MONTHS OR EARLIER IF OCCLUSION/ MIGRATION OCCURRED AND THAT THE MEDIAN PATENCY PERIOD OF INSIDE STENT WAS 222 DAYS (7.4 MONTHS) AND OUTSIDE STENT WAS 99 DAYS(3.3 MONTHS), THE ADVERSE EVENTS NOTED WITH BOTH ¿INSIDE¿ AND ¿OUTSIDE¿ STENTS OF MIGRATION, OCCLUSION, STONE FORMATION, PANCREATITIS AND CHOLANGITIS WILL BE DEALT WITH IN SEPARATE FILES (B)(4). THIS FILE WILL DEAL THE CASES OF PANCREATITIS WITH THE USE OF THE ¿OUTSIDE STENT¿. THIS PANCREATITIS HAS BEEN CONFIRMED AS LINKED TO THE STENT EXCEEDING THE MAXIMUM 3 MONTH INDWELL PERIOD USER ERROR BY CLINICAL ADVISEMENT. EARLY COMPLICATIONS OF PANCREATITIS WERE NOTED WITH BOTH THE INSIDE AND OUTSIDE STENTS. AS THE INFORMATION REPORTED NOTES 6 SESSIONS LEADING TO PANCREATITIS AND ONLY FIGURES OF 3 AND 1 MENTIONED WITH THE INSIDE AND OUTSIDE STENT RESPECTFULLY FOR THE PURPOSE OF THIS INVESTIGATION A WORST CASE APPROACH WAS TAKEN OF 3 STENT FOR INSIDE AND 3 FOR OUTSIDE LEADING TO PANCREATITIS. IT WAS NOTED THAT ALL PATIENTS IMPROVED WITH CONSERVATIVE TREATMENT. THERE WERE NO SEVERE ADVERSE EVENTS. A DEFINITIVE ROOT CAUSE CAN BE ATTRIBUTED TO THE USER ERROR OF THE IFU NOT BEEN FOLLOWED WITH THE DEVICE EXCEEDING THE MAXIMUM 3 MONTH INDWELL PERIOD IN THE PATIENT. WHEN THE DEVICE IS USED OUTSIDE ITS STATED PARAMETERS IT MAY LEAD TO OUTCOMES THAT WERE NEVER INTENDED TO HAPPEN AND WERE NEVER STUDIED. IT CAN BE NOTED THAT AS PER THE INSTRUCTIONS FOR USE, PANCREATITIS IS A KNOWN POTENTIAL COMPLICATION. COMPLAINT IS CONFIRMED BASED ON CUSTOMER TESTIMONY. ACCORDING TO THE INFORMATION REPORTED, BILIARY STRICTURES WERE SUCCESSFULLY TREATED IN 14 PATIENTS WERE STENTING WAS POSSIBLE. BILIARY STRICTURES IMPROVED IN 7 PATIENTS ON REPEAT ERC, WHO THEREFORE HAD THEIR STENTS REMOVED. STENT REPLACEMENT WAS PERFORMED EVERY 6 TO 12 MONTHS FOR THE OTHER 7 PATIENTS. IT WAS NOTED THAT ALL PATIENTS IMPROVED WITH CONSERVATIVE TREATMENT. THERE WERE NO SEVERE ADVERSE EVENTS. COMPLAINTS OF THIS NATURE WILL CONTINUE TO BE MONITORED FOR POTENTIAL EMERGING TRENDS.

Additional Manufacturer Narrative · 0

THE EXACT RPN IS UNKNOWN BUT BASED ON THE DEVICE NAME OF COTTON-LEUNG BILIARY STENT, THE TWO POSSIBLE 510K NUMBERS ARE K851962 / K851962. INVESTIGATION IS STILL PENDING. A FOLLOW UP MDR WILL BE SUBMITTED TO INCLUDE THE INVESTIGATION CONCLUSIONS.

Description of Event or Problem · 0

CLINICAL INPUT WAS RECEIVED ON THE 02-FEB-2020, CONFIRMING THAT THE PANCREATITIS EXPERIENCED BY THE PATIENTS WITH THE OUTSIDE STENT WILL BE CONSIDERED AS USER ERROR FOR EXCEEDING THE MAXIMUM INDWELL PERIOD OF 3 MONTHS FOR STENT. SUPPLEMENT REPORT IS BEING SUBMITTED TO CAPTURE THE PRODUCT PROBLEM CODE.

Description of Event or Problem · 0

SUPPLEMENTAL REPORT IS BEING SUBMITTED DUE TO THE COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 0

CHANGES WERE MADE TO THE FILE ON THE 26-JAN-2021, THIS FILE IS NOW CAPTURING 3 CASES OF PANCREATITIS RELATED TO THE OUTSIDE STENT WHICH WAS USED ON LABEL. MDR REF # 3001845648-2020-00547 IS CAPTURING THE OTHER PANCREATITIS CASES USED OFF LABEL. THIS UPDATE IS TO ALIGN WITH COOK IRELAND'S COMPLAINT PROCESS, NO NEW INFORMATION WAS RECEIVED. EVENT DESCRIPTION HAS BEEN UPDATED. KOIZUMI ET AL. 2020 ¿ENDOSCOPIC STENT PLACEMENT ABOVE THE SPHINCTER OF ODDI FOR BILIARY STRICTURES AFTER LIVING DONOR LIVER TRANSPLANTATION¿. SIXTEEN CONSECUTIVE PATIENTS WITH A MEDIAN AGE OF 55.5 YEARS (RANGE, 36¿67 YEARS), SUFFERING FROM BILIARY STRICTURES AFTER LDLT WITH DUCT-TO-DUCT BILIARY RECONSTRUCTION FOR END-STAGE LIVER DISEASE OR ACUTE LIVER FAILURE AT OUR HOSPITAL BETWEEN 2004 AND 2016 WERE ELIGIBLE. ANASTOMOTIC BILIARY STRICTURES WERE DIAGNOSED BY ENDOSCOPIC RETROGRADE CHOLANGIOGRAPHY (ERC). OUTSIDE STENTS WERE PLACED ACROSS THE SPHINCTER OF ODDI, WITH THEIR DISTAL END EXPOSED INTO THE DUODENUM. INSIDE STENTS BEING PLACED ABOVE THE SPHINCTER OF ODDI (FIG. 1B, E). ENDOSCOPIC SPHINCTEROTOMY (EST) WAS NOT PERFORMED FOR BOTH METHODS AT INITIAL ATTEMPT. THE SHAPE OF A 7-10FR PLASTIC STENT FLEXIMA; BOSTON SCIENTIFIC JAPAN, TOKYO, JAPAN. COTTON-LEUNG SOF-FLEX BILIARY STENT; COOK JAPAN, JAPAN. THROUGHPASS; GADELIUS MEDICAL, TOKYO, JAPAN) WAS MODIFIED AS AN INSIDE STENT. THE DISTAL FLAP OF THE STENT WAS REMOVED BUT INSTEAD A NYLON THREAD WAS ATTACHED TO THE DISTAL SIDE HOLE BEYOND THE DUODENAL PAPILLA TO FACILITATE RETRIEVAL OF THE INSIDE STENT, AS PREVIOUSLY REPORTED. IN THIS STUDY, WE ANALYZED A TOTAL OF 98 ERC SESSIONS FOR 16 PATIENTS UNDERGOING ENDOSCOPIC TREATMENT OF BILIARY STRICTURE AFTER LDLT WITH DUCT-TO-DUCT BILIARY RECONSTRUCTION. BILIARY STENTS WERE PLACED IN 87 SESSIONS: 77 INSIDE SESSIONS AND 10 OUTSIDE SESSIONS. EARLY COMPLICATIONS WERE PANCREATITIS (6 SESSIONS) ALL IMPROVED WITH CONSERVATIVE TREATMENT. AS PER INFORMATION SUPPLIED IN THE PAPER 1 INSIDE STENT AND 3 OUTSIDE LEAD TO PANCREATITIS. THIS FILE INVOLVES A WORST-CASE APPROACH OF 3 CASES OF PANCREATITIS WITH THE OUTSIDE STENT.

Additional Manufacturer Narrative · 1

THE EXACT RPN IS UNKNOWN BUT BASED ON THE DEVICE NAME OF COTTON-LEUNG BILIARY STENT, THE TWO POSSIBLE 510K NUMBERS ARE K851962 / K851962. INVESTIGATION IS STILL PENDING. A FOLLOW UP MDR WILL BE SUBMITTED TO INCLUDE THE INVESTIGATION CONCLUSIONS.

Description of Event or Problem · 1

KOIZUMI ET AL. 2020 ¿ENDOSCOPIC STENT PLACEMENT ABOVE THE SPHINCTER OF ODDI FOR BILIARY STRICTURES AFTER LIVING DONOR LIVER TRANSPLANTATION¿. SIXTEEN CONSECUTIVE PATIENTS WITH A MEDIAN AGE OF 55.5 YEARS (RANGE, 36¿67 YEARS) SUFFERING FROM BILIARY STRICTURES AFTER LDLT WITH DUCT-TO-DUCT BILIARY RECONSTRUCTION FOR END-STAGE LIVER DISEASE OR ACUTE LIVER FAILURE AT OUR HOSPITAL BETWEEN 2004 AND 2016 WERE ELIGIBLE. ANASTOMOTIC BILIARY STRICTURES WERE DIAGNOSED BY ENDOSCOPIC RETROGRADE CHOLANGIOGRAPHY (ERC). OUTSIDE STENTS WERE PLACED ACROSS THE SPHINCTER OF ODDI, WITH THEIR DISTAL END EXPOSED INTO THE DUODENUM. INSIDE STENTS BEING PLACED ABOVE THE SPHINCTER OF ODDI (FIG. 1B, E). ENDOSCOPIC SPHINCTEROTOMY (EST) WAS NOT PERFORMED FOR BOTH METHODS AT INITIAL ATTEMPT. THE SHAPE OF A 7-10FR PLASTIC STENT FLEXIMA; BOSTON SCIENTIFIC JAPAN, TOKYO, JAPAN. COTTON-LEUNG SOF-FLEX BILIARY STENT; COOK JAPAN, JAPAN. THROUGHPASS; GADELIUS MEDICAL, TOKYO, JAPAN) WAS MODIFIED AS AN INSIDE STENT. THE DISTAL FLAP OF THE STENT WAS REMOVED BUT INSTEAD A NYLON THREAD WAS ATTACHED TO THE DISTAL SIDE HOLE BEYOND THE DUODENAL PAPILLA TO FACILITATE RETRIEVAL OF THE INSIDE STENT, AS PREVIOUSLY REPORTED. IN THIS STUDY, WE ANALYZED A TOTAL OF 98 ERC SESSIONS FOR 16 PATIENTS UNDERGOING ENDOSCOPIC TREATMENT OF BILIARY STRICTURE AFTER LDLT WITH DUCT-TO-DUCT BILIARY RECONSTRUCTION. BILIARY STENTS WERE PLACED IN 87 SESSIONS: 77 INSIDE SESSIONS AND 10 OUTSIDE SESSIONS. EARLY COMPLICATIONS WERE PANCREATITIS (6 SESSIONS) ALL IMPROVED WITH CONSERVATIVE TREATMENT. IT IS UNKNOWN WHETHER THE PANCREATITIS OCCURRED WITH THE INSIDE OR OUTSIDE STENT, THIS FILE IS COVERING THE POTENTIAL OF BOTH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
961764 COTTON-LEUNG BILIARY STENT FGE CATHETER, BILIARY, DIAGNOSTIC FGE COOK IRELAND LTD UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 56 YR Unknown Required Intervention