SYRINGE 50ML LL TIP 1ML
Report
- Report Number
- 1911916-2020-00846
- Event Type
- Malfunction
- Date Received
- September 3, 2020
- Date of Event
- August 15, 2020
- Report Date
- August 19, 2020
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- FMF
- UDI-DI
- 30382903096535
- PMA / PMN Number
- K980987
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE CUSTOMER'S ADDRESS IS UNKNOWN. (B)(6) HAS BEEN USED AS A DEFAULT. (B)(4). INVESTIGATION SUMMARY: A DEVICE HISTORY RECORD REVIEW WAS PERFORMED FOR PROVIDED LOT NUMBER 1760575. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS REPORTED INCIDENT. TO AID IN THE INVESTIGATION OF THIS INCIDENT, 2 PICTURE SAMPLES WERE RECEIVED FOR EVALUATION BY OUR QUALITY TEAM. THROUGH EXAMINATION OF THE RETURNED SAMPLES, THE SYRINGE HAS SCALE MARKINGS MISSING. INVESTIGATION CONCLUSION: OUR QUALITY TEAM WILL CONTINUE TO MONITOR THE MANUFACTURING PROCESS FOR THIS DEFECT AND OTHER EMERGING TRENDS. ROOT CAUSE DESCRIPTION: THE CAUSE OF THIS DEFECT COULD HAVE RESULTED DURING THE SYRINGE ASSEMBLY PRINTING PROCESS WHERE THE MACHINE RAN OUT OF INK. THIS WOULD RESULT IN NO MARKINGS PRINTING ON THE SYRINGE, AND THE DEFECT WENT UNDETECTED DURING THE NEXT PROCESS. RATIONALE: FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME.
IT WAS REPORTED THAT SYRINGE 50 ML LL TIP 1 ML WAS MISSING SCALE MARKINGS. THIS WAS DISCOVERED ON 4 OCCASIONS BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 309653, BATCH NO: 0062316. IT WAS REPORTED THAT: SYRINGE WAS MISSING INCREMENT ML MARKINGS. EVENT DESCRIPTION PER EMAIL STATES: VERBIAGE RECEIVED, I FOUND 3 OF THESE WITH NO LABEL PRINTED ON THEM. I LEFT THIS ONE IN THE PACKAGING. ALL FROM THE SAME LOT NUMBER. THE REST OF THE SYRINGES FROM THIS LOT ALL APPEAR TO BE FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 951163 | SYRINGE 50ML LL TIP 1ML | PISTON SYRINGE | FMF | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 309653 | 0062316 | 30382903096535 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |