FDA Adverse Event Malfunction Summary report: N

BARREL 3ML LL WWD WITH SIL NO LOGO BNS

MDR report key: 10488064 · Received September 3, 2020

Report

Report Number
1213809-2020-00597
Event Type
Malfunction
Date Received
September 3, 2020
Date of Event
August 12, 2020
Report Date
September 9, 2020
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY THREE PHOTOS AND SIX LOOSE 3ML BARRELS WERE RECEIVED AND EVALUATED. IT WAS OBSERVED IN THE PHOTOS AND PHYSICAL SAMPLES, THERE WAS A SMALL PERPENDICULAR SCRATCH PRESENT OUTSIDE THE GRAD LINES NEAR THE 0.5ML MARKING. THE SCRATCHES WERE APPROXIMATELY 0.25" IN LENGTH. ON ALL THE SAMPLES, THE SCRATCH APPEARED TO BE COSMETIC IN NATURE WITH NO EFFECT TO THE FORM FIT OR FUNCTION OF THE DEVICE, WHICH WAS ACCEPTABLE PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE SCRATCHED BARREL DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. IT WAS LIKELY DUE TO A BARREL CAUGHT IN THE GUIDE AND CONTACTING THE PRODUCT AS IT PASSED BY. SINCE THE DEFECTS OBSERVED ARE WITHIN THE ACCEPTABLE LIMITS, NO CORRECTIVE ACTIONS ARE NECESSARY. A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 9211905.

Description of Event or Problem · 0

IT WAS REPORTED THAT (B)(4) BARREL 3ML LL WWD WITH SIL NO LOGO BNS WERE DAMAGED, BUT STILL OPERABLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "IT WAS REPORTED THAT (B)(4) SYRINGES HAVE SCRACTHES/SLITS ON THE OUTSIDE WALL OF THE BARREL. RELATED TW# (B)(6) PER EMAIL: AGAIN, WE FOUND MORE UNITS WITH THIS CONDITION AND PRODUCTION HAS REQUESTED TO RETAIN ALL THE MATERIAL COMING FROM THIS LOT: 9211905. WE HAVE A TOTAL OF (B)(4)EAS."

Additional Manufacturer Narrative · 1

MEDICAL DEVICE EXPIRATION DATE: NA. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT 111172 BARREL 3ML LL WWD WITH SIL NO LOGO BNS WERE DAMAGED, BUT STILL OPERABLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "IT WAS REPORTED THAT (B)(4) SYRINGES HAVE SCRATCHES/SLITS ON THE OUTSIDE WALL OF THE BARREL. RELATED TW# (B)(4). PER EMAIL: AGAIN, WE FOUND MORE UNITS WITH THIS CONDITION AND PRODUCTION HAS REQUESTED TO RETAIN ALL THE MATERIAL COMING FROM THIS LOT: 9211905. WE HAVE A TOTAL OF (B)(4) EAS."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
950991 BARREL 3ML LL WWD WITH SIL NO LOGO BNS PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 9211905

Patients

Seq Age Sex Outcome Treatment
1 Other