CUP: VERSAFITCUP ACETABULAR SHELL 58
Report
- Report Number
- 3005180920-2020-00576
- Event Type
- Injury
- Date Received
- September 3, 2020
- Date of Event
- August 4, 2020
- Report Date
- December 3, 2020
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- MEH
- UDI-DI
- 07630030808203
- PMA / PMN Number
- K083116
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
VISUAL INSPECTION: IT IS POSSIBLE THAT A SIZING OCCURRED DUE TO AN INCORRECT USAGE: - WRONG SCREWING OF THE SHELL VERSAFICUP #58 DM REF.01.26.10.0139 WITH THE IMPACTING RING #58 REF.01.26.10.0139
BATCH REVIEW PERFORMED BY MEDACTA REGULATORY AFFAIRS DEPARTMENT ON 13.08.2020: LOT 160207: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 2-MAY-2016. EXPIRATION DATE: 2021-04-18. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT. OTHER DEVICE INVOLVED IN THE EVENT: CUP: VERSAFITCUP CC TRIO 01.26.45.1158 ACETABULAR SHELL CC TRIO NO-HOLE Ø 58 LOT. 2000144 (K122911). BATCH REVIEW PERFORMED BY MEDACTA REGULATORY AFFAIRS DEPARTMENT ON 13.08.2020: LOT 2000144 : (B)(4) ITEMS MANUFACTURED AND RELEASED ON 30-APR-2020. EXPIRATION DATE: 2025-04-20. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT.
THE SURGEON WANTED TO IMPLANT A VERSAFITCUP DM S58 AND AFTER HE HAD A PRESS-FIT THE IMPACTING RING DID NOT COME OFF. BY TRYING SEVERAL TIMES AFTERWARD ALSO WITH A VERSAFITCUP CC TRIO HE HAD TO CEMENT A NEW ONE- APRICOT. THE BONE STRUCTURE WAS DAMAGED. THE SURGEON CEMENTED A NEW CUP ON THE BONE DAMAGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 954936 | CUP: VERSAFITCUP ACETABULAR SHELL 58 | ACETABULAR SHELL | MEH | MEDACTA INTERNATIONAL SA | 01.26.58MB | 160207 | 07630030808203 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |